Effluent Decontamination Veterinary Autoclave Bulk Steam
Sterilization For Animal Labs
Specification
Model | Chamber Size | Overall Size | Net Weight | Steam | Water | Power |
W*H*D mm | W*H*D mm | Kg | Kg/cycle | Kg/cycle | KVA | |
MD-0.11V | 400*400*700 | 1000*1800*1040 | 650 | 14 | 15 | 2 |
MD-0.16V | 400*400*1000 | 1000*1800*1340 | 700 | 16 | 20 | 2 |
MD-0.20V | 400*400*1300 | 1000*1800*1640 | 750 | 18 | 25 | 2 |
MD-0.24V | 600*600*670 | 1200*2000*1010 | 800 | 20 | 30 | 2 |
MD-0.30V | 600*600*840 | 1200*2000*1180 | 850 | 22 | 35 | 2 |
MD-0.36V | 600*600*1000 | 1200*2000*1340 | 900 | 25 | 40 | 2 |
MD-0.45V | 600*600*1250 | 1200*2000*1590 | 980 | 27 | 50 | 2 |
MD-0.60V | 600*600*1670 | 1200*2000*2010 | 1100 | 30 | 60 | 3 |
Program Parameter
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test programs | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leak Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |
Features
Description
Feature of the veterinary autoclave is the control of the
sterilization phase according to a pressure signal, rather than temperature. Obviously, the
temperature readout varies as a function of the position of the sensor, whilst the pressure is
uniform throughout the chamber.
However, process control is performed by reading in the chamber (and from inside the product) temperature signals whose match with
the pressure value, as constantly verified by the controller, demonstrates that the
autoclave is operating correctly.
This principle, in saturated-steam conditions (when there is a
two-way T/P correspondence) allows to obtain uniformities in sterilization temperature which
are far better than the currently required level of ±1.0 °C.
Applications
Competitive Advantage
Effluent Decontamination.
Available on gravity and pre-vacuum units. Retains liquid effluent,
if any, within the chamber for the duration of the sterilization cycle. Aerosol effluent
passes through a 0.2 micron hydrophobic HEPA filter with >99.999% efficiency; the filter is
located within the chamber and is sterilized in situ to eliminate the need for disposal after
each cycle. The filter is easily replaced by the user when required without special tools or
service personnel.