PLC Control Systems Horizontal Autoclave Sterilizer With Advanced Sterilization Cycles
Specification
This Horizontal Autoclave is designed and manufactured specially for laboratories, hospital CSSD, clinic, operating room to sterilize experiment tools, medical instruments, sterile clothing, dressings, vessels, apparatuses, operating tools, culture medium etc.
Designed and manufactured according to ISO 9001, ISO 13485 system and latest moist steam sterilizer related standards, which has the obvious advantage of low failure rate and easily operation to ensure your satisfied using smoothly.
Control system consists of Mitsubishi PLC, 65526 colors TFT touch Screen, No paper recorder and micro printer etc to ensure safe and accurate operation.
Multiple sterilization programs (15 sets sterilization cycles, 2 sets test cycles) with icon operation for user selection, all parameters can be reset by user according to certain requirements.
Sterilization cycle can be start by one touch key and runs automatically till cycle finished, the main parameters of the cycles, like pressure, temperature etc. will be recorded into no paper recorder and printed by micro printer.
History temperature data in the recorder can be displayed on the screen or restore in the USB flash disks.
Features
Parameters:
Model | MD-0.24V |
Door type | Vertical sliding single/double door |
Max working pressure | 0.23Mpa |
Working vacuum degree | <-0.086Mpa |
Max working temperature | 134℃ |
Chamber size | 1500*400*400 mm(L*W*H) |
External dimension | 1850*1250*1800mm(L*W*H) |
Net weight | 800Kg |
Utilities | |
Electric power | AC380V, 50Hz, 3Ph/N/E, 3KVA+18KW(for built-in steam generator) |
Water | 0.15----0.3MPa, Hardness <50ppm/pH Neutral |
Compressed air | 0.5~0.7Mpa |
Sterilization cycles
1, B-D Test
3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min.
2, Vacuum Leakage Test
Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect.
Vacuum degree: -92Kpa, Holding time: 900 seconds, Max pressure rise: <=2Kpa
3, Instruments/Tools
3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min.
4, Clothing
3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min.
5, Liquid
15 min exchange, expose temperature 121℃, expose time 15 min, cooling time 25 min.
6, User Programmable
All preset parameters can be resettable; total 15 kinds programs can be customized
7, Main adjustable parameters:
Sterilization temperature rage: at 105°C to 134°C
Sterilization time rage: 0---99 minutes
Vacuum times: 0---9 times
Drying time: 0---99 minutes