Incorporated Air Filter ETO Sterilization Machine For Ethylene Oxide Gas Sterilization

Brand Name:Medpharm
Certification:ISO
Model Number:SQ-H series
Minimum Order Quantity:1 unit
Place of Origin:China
Supply Ability:50 units per month
Contact Now

Add to Cart

Verified Supplier
Address: B631, HUAFANG INTERNATIONAL MANSION, No.178 CHENGBEI STR., ZHANGJIAGANG, JIANGSU, PRC, 215600
Supplier`s last login times: within 48 hours
Product Details Company Profile
Product Details

EO PROCESSING STEPS

  • Preconditioning/conditioning
    • Exposure to RH and temperature
    • Ensure uniformity of these conditions
  • Sterilization cycle
    • Exposure to EO gas
  • Aeration
    • Dissipation of remaining gases

DECISIVE PROCESS PARAMETERS

  • Gas concentration >400mg/L
  • Temperature ~100 – 140ºC
  • Relative humidity ~35 – 80%
  • Exposure (dwell) time 2 – 10 hours

DEEP VACUUM CYCLE

SHALLOW VACUUM CYCLE

FACTORS AFFECTING CYCLE SUCCESS

  • Bioburden
  • Product/package properties
  • Loading configuration
  • Cycle parameters

DETERMINE THE STANDARD

  • AMI/ISO 11135-01 4ed

“Sterilization of health care products – Ethylene oxide - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices”

  • Europe – EN 550

EO VALIDATION OVERVIEW

  • Process development
  • Product compatibility
  • Commissioning
  • PQ – Physical
  • PQ – Microbiological
  • Certification
  • Revalidation

PROCESS CONTROL

  • Must assure that validated process parameters are met
    • Temperature
    • RH
    • Gas concentration
  • Biological indicators are used to demonstrate lethality
  • Microprocessors are used to control process

RELEASE MECHANISMS

  • Documentation showing that processing specification are met
  • Successful results of tests
    • Sterility of BI
    • EO residues
    • Packaging
    • Pyrogens

PARAMETRIC RELEASE

  • BIs not used in release
  • Validation more involved
  • Routine control more rigorous
  • AAMI TIR20:2001 “Parametric release for ethylene oxide sterilization”

PRODUCT COMPATIBILITY

  • Post sterilization testing for
    • Device functionality
    • Package integrity and strength
    • Residue dissipation rates
    • Impact of re-sterilization

COMMISSIONING

  • Equipment specifications/diagram
  • Calibration records
  • Profiles for
    • Preconditioning (temp. and RH)
    • Aeration rooms (temp.)
    • Empty chamber temperature distribution

ModelChamber size(L*W*H)cmOverall size(L*W*H)cm
SQ-H8055×27×3874×47×92
SQ-H12060×40×5077×61×106
SQ-H22074×50×6092×68×115
SQ-H36080×65×7097×84×131
SQ-H46080×72×8098×92×148
SQ-H60088×80×86116×100×155
SQ-H80095×90×95115×110×166