Autoclave Laboratorio Hospital Large Steam Sterilizer Sterilaztion
Design pressure | 0.29MPa |
Rated working pressure | 0.21Mpa |
Design temperature | 150 ℃ |
Rated working temperature | 121 ℃ / 134 ℃ |
Voltage | 220V |
Specifications | SQ-Z |
Power | 16.0KW |
Temperature display | LCD screen |
Heating method | Electric |
Features:
1. All of the B class program. flash program, cleaning program,
user defined program.
2. LCD Display
3. The parameter of pregram is adjustable
4. Vaccum Test/BD Test.
5. Helix test passed
6. The Door for protecting open.
7. Quickly sterilize
8. Can keep temperature for next quickly
sterilizing.
9. Easy to maintain
10. Easy to operate
Paramter:
Power: 2.4 KW 220V/50HZ ( Can Supply other Power)
Product Size(L×W×H):580mm x 440mm x 360mm Volume: 8L
Chamber Size: Φ170mm×320mm
Distance between foot pads of product(L×W): 480mm x 320mm
1 General Introduction
Steam Sterilizer is operated by doctors or professionals and is
designed specially for clinic, hospital, laboratory etc. The
sterilizer uses microprocessor with intelligence control system,
and humanistic interface, operate easily, safety and reliably. The
parameters and conditions of the sterilizer will be displayed on
the digital screen during the processing. For ensuring the
reliability of sterilization, the machine will do trouble
self-diagnose and self-protect automatically during overheat or
overpressure situations. Inside of the sterilizer has a collector
of condensate water that
prevents the steam from polluting the environment.
2 Using Range
This sterilizer is for sterilization of invasive medical devices.
It can prevent cross infection.. This sterilizer is highly
penetrability for department of hospital, stomatology,
ophthalmology, and biological research institute. Sterilizing
surgical equipments, stomatology instruments and
syringes etc. It can sterilize the wrapped or non-wrapped, solid,
hollow load products type A and porous products as represented by
the test loads in the standard “EN13060:2004+A2:2010”. This
sterilizer is also can use in non-medical and veterinary
applications.
3 Parameters and process of the sterilizer
The Tank’s capacity :
The max volume of waste water tank: 4L
The max volume of clean water tank: 4L