GF333B three phase portable reference meter test equipment with 128st harmonics measurement


GF333B three phase portable reference meter is a reference standard with the characteristics of wide-range, multifunction and high-precision. It has accuracy class of 0.02. By adopting techniques of DSP, embedded system and automatic temperature balancing and other compensation, it has not only advantages of light weight, small size, high precision, high performance, powerful function, original interface, ease-to-use, but also work stability.The product can be widely used in fields of electric energy measuring, electric energy laboratories and other relevant industry, not only in laboratories but also at the industrial field.

FEATURES

  • High accuracy up to 0.05%
  • Wide testing range: voltage 5V-480V, current 1mA-120A
  • Testing type: 3 phase 4 wire, 3 phase 3 wire
  • Testing mode: active power, reactive power, apparent power
  • Harmonic analysis function
  • Waveform display function
  • Vector diagram function
  • Energy accumulating function
  • Internal pulse input port, testing meter’s error directly

PARAMETERS

Electrical parameters 
Accuracy0.05%
Voltage supply220V±10% or 110V±10%, 50/60Hz
Voltage measurement 
Range5V-480V
Error±0.02% (50V-480V), ±0.05% (5V-50V)
Display range5.000000V-480.0000V
Harmonic2nd-63st
Current measurement 
Range1mA-120A
Error±0.02% (0.2A-120A), ±0.05% (1mA-0.2A)
Display range1.000000mA-120.0000A
Power measurement 
Active power±0.02% (0.2A-120A)
±0.05% (0.01A-0.2A)
±0.1% (0.001A-0.01A)
Reactive power±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A)
Apparent power±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A)
Energy error 
Active energy±0.02% (0.2A-120A)
±0.05% (0.01A-0.2A)
±0.1% (0.001A-0.01A)
Reactive energy±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A)
Apparent energy±0.05% (0.2A-120A), ±0.1% (0.001A-0.2A)
Phase measurement 
Range0°-360°
Error±0.005°
Display range0.0001°-359.99°
Frequency measurement 
Range45-65Hz
Display range45.0000-65.0000
Accuracy0.002Hz
Power Factor measurement 
Range-1.0 ~ 0 ~ +1.0
Resolution0.0001
Accuracy0.0005
Energy pulse 
High frequency output(CH)12000Hz
Low frequency output(CL)5000Hz
Pulse ratio1:1
Output level5V
Input level5V
Input frequencyMax. 1MHz
Mechanical parameters 
Dimensions (W×H×D) (mm)480×138×486
Weight (kg)10
Environmental conditions 
Ambient temperature-10°C to 45°C
Influence of external fields≤0.05 %/mT
Relative humidity10%-80%