GF6018A 20A 1050V Multifunction Electric Ammeter and Voltmeter calibrator for clamp meter calibration
The equipment of clamp type multimeter calibrator is compliance
with national verification regulation: JJG124-2005 “Ammeter,
voltmeter, power meter and resistance meter verification rules” and
related national standards. 0.05 single-phase AC/DC standard
source, can verify magnitude 0.2 and the following AC/DC voltmeter,
ammeter, frequency meter, resistance meter. The source signals by
using DSP and 16-bit high-speed analog-to-digital converters of
controllable sine wave, the distorted wave signal source.The
equipment comes with RS-232 interface, can be connected with PC to
a semi-automatic verification system. It has precise interface and
multi-functions including verification, storage and query.The
equipment adopts color LCD screen, chart character display is
clear, with high precision, stable and reliable, convenient
operation and flexible characteristics.
Calibration of Voltmeter, Ammeter & Wattmeter using
Potentiometer The calibration is the process of checking the
accuracy of the result by comparing it with the standard value. In
other words, calibration checks the correctness of the instrument
by comparing it with the reference standard.
Calibration unit has a Voltmeter and an Ammeter, which can be
Calibrated by Potentiometer. A Potentiometer is an instrument
designed to measure an unknown voltage by comparing it with a known
voltage. It has a high degree of accuracy.
The best method of calibration is to measure the true value of
current or voltage in the circuit and compare it with the value
measured by the given ammeter or voltmeter. The true value of
current or voltage in the circuit can be measured up to a fair
grade of accuracy by using a potentiometer.
Theory
Calibration of multimeter in measurement technology and metrology
is the comparison of measurement values delivered by a device under
test with those of a calibration standard of known accuracy. Such a
standard could be another measurement device of known accuracy, a
device generating the quantity to be measured such as a voltage, or
a physical artefact, such as a metre ruler.
The outcome of the comparison can result in no significant error
being noted on the device under test, a significant error being
noted but no adjustment made, or an adjustment made to correct the
error to an acceptable level. Strictly speaking, the term
calibration means just the act of comparison, and does not include
any subsequent adjustment.
The calibration standard of multimeter and clamp meter is normally
traceable to a national standard held by a National Metrological
Institute.
FEATURES
It can be semi-automatic or manual inspection of analog multi meter, clamp type current meter, all kinds of electric meter, (voltmeter, ammeter, frequency meters, resistance meter) basic error
The equipment can be generated with 2-31 harmonic wave, harmonic number, order, amplitude and harmonic phase of fundamental wave can be programmed
Power amplifier working frequency range is 40Hz-1KHZ, has a good linear. Current amplifier is constant current source; voltage amplifier is constant voltage source. Because of light weight, this device is more suitable for field use
With RS-232 interface, this device is controlled by PC through the PC software (optional), can be automatic or manual inspection, and the results are processing and management
Has a non-volatile memory, it can store 500 pieces of the raw data of tested meter, for access and upload
PARAMETERS
Electrical parameters | ||
---|---|---|
Power supply | Single phase AC 220V±10% or 110V±10%, 50/60HzAC | |
AC Voltage output | ||
range | 200mV, 400mV, 2000mV, 4000mV, | |
Range 20V-1000V | Accuracy | ±(0.03%RD+0.02%FS) |
Stability | ≤0.01% FS /60s | |
Range 200mV-4000mV | Accuracy | ±(0.3%RD+0.2%FS) |
Stability | ≤0.04% FS /60s | |
Load capacity | 20VA | |
Waveform distortion | ≤0.3% | |
Adjustable Range | 0-120% FS (1000V not included) | |
Fineness | 5×10-5 | |
AC Current Output | ||
Range | 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A | |
Range 0.2A-20A | Accuracy | ±(0.03%RD+0.02%FS) |
Stability | ≤0.01% FS /60s | |
Range 2mA-40mA | Accuracy | ±(0.3%RD+0.2%FS) |
Stability | ≤0.04% FS /60s | |
Load capacity | 20VA | |
Waveform distortion | ≤0.3% | |
Adjustable Range | 0-120% FS | |
Fineness | 5×10-5 | |
Frequency Output | ||
Range | 45-65Hz | |
Fineness | 0.001Hz | |
Accuracy | 0.01Hz | |
Harmonic Wave | ||
Times | 2-31 | |
Extent | 0-20% | |
Each harmonic phase fineness | 0.01·N (N is harmonic order) | |
DC voltage output | ||
Range | 200mV, 400mV, 2000mV, 4000mV, | |
Load capacity | 20W | |
Adjustable range | 0-110% FS (1000V not included) | |
Adjustable fineness | 5×10-5 | |
Stability | ≤0.01% FS /60s (peak-peak value) | |
Accuracy | ±(0.03%RD+0.02%FS) | |
Ripple wave | ≤1% FS | |
DC current output | ||
Range | 20uA, 40uA, 200uA, 400uA, 2mA, 4mA, | |
Load capacity | 20W | |
Adjustable Range | 0-110% FS (1000V not included) | |
Adjustable Fineness | 5×10-5 | |
Stability | ≤0.01% FS /60s (peak-peak value) | |
Accuracy | ±(0.06%RD+0.04%FS) | |
Ripple wave | ≤1% FS | |
Resistance box | ||
Range | 0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ, | |
Fineness | 0.02% | |
Range 0-2kΩ, 0-20kΩ, 0-200kΩ | ||
Range 0-2kΩ | Adjustable Range | 0-100%FS |
Accuracy | ±(0.12%RD+0.08%FS) | |
Range 0-200Ω,0-2MΩ | Adjustable Range | 0-100% FS |
Accuracy | ±(0.3%RD+0.2%FS) | |
Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ | ||
Range 2kΩ-4kΩ | Adjustable Range | 5%-100%FS |
Accuracy | ±(0.12%RD+0.08%FS) | |
Range 200-400Ω, | Adjustable Range | 5%-100% FS |
Accuracy | ±(0.3%RD+0.2%FS) | |
Clamp meter range | ||
200A | Accuracy | ±0.2% |
600A | Accuracy | ±0.5% |
1000A | Accuracy | ±1% |
Dot frequency square wave output | ||
Range | 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz | |
Square wave amplitude | 4.5V | |
Setting accuracy | 5×10-5 | |
Mechanical parameters | ||
Dimension (L×W×H) (mm) | 440x360x160 | |
Weight (kg) | 12 | |
Environmental conditions | ||
Operating temperature | 20°C±10°C | |
Relative humidity | ≤85%RH |