Add to Cart
Contant Temperature Humidity Test Chamber 225 Liters Lab Testing Equipment
Test Chamer Application:
This 225 Liters constant temperature humidity test chamber for lab is mainly used for products in accordance with the requirements of the national standard or user-defined requirements, at low temperatures, high temperature, humidity and temperature cycling conditions, physical and other relevant characteristics of the product environmental simulation test, after tested by detecting to judge the performance of the product, it is still able to meet predetermined requirements in order for the product design, improvement, identification and factory inspection use.
Temperature Humidity Test Chamber 225 Liters Deatil Specification
Model No. | PT-2091A | PT-2091B | PT-2091C | PT-2091D | PT-2091E PT-2091F |
Product Name | Contant Temperature Humidity Test Chamber 225 Liters Lab Testing Equipment | ||||
Inner Cabinet (cm) | 40×40×50 | 50×50×60 | 50×60×75 | 60×80×85 | E:100×80×100 F: 100×100×100 |
Temperature Range (℃) | F: -20℃, L: -40℃, T: -60℃, G: -70℃ ~ 150℃ | ||||
Humidity Range (%RH) | 20%R.H ~ 98%R.H | ||||
Temperature Fluctuation (℃) | ±0.5℃ | ||||
Temperature uniformity (℃) | ≤ ±2℃ | ||||
Humidity fluctuation (%RH) | ±2.5%RH | ||||
Humidity uniformity (%RH) |
| ||||
Heating rate (℃) | 3℃/min (No-load) | ||||
Cooling rate (℃) | 0.7 ~ 1℃/min (No-load) | ||||
Temperature and humidity control system | Imported LCD digital display controller. | ||||
Heated Humidification system | Nichrome wire-efficiency heater, UL-shaped stainless steel plating heater | ||||
Cooling System |
| ||||
Inner Material | SUS#304 stainless steel. | ||||
Outer Material | Spray color galvanized steel sheet. | ||||
Insulation Materials | Ultra-fine glass fiber insulation cotton. | ||||
Communication | Equipped with a standard RS232C / RS485 communication interface, the maximum transmission distance of 1.2Km, the fastest communications up to 9600bps, can be realized with a computer connected to the remote monitoring, easy to operate. | ||||
Subsidiary function | Fault alarm and reason prompts, you can change the name of the fault, the upper temperature limit protection, timing and the reservation (automatic stop and start function), multiple (up to 6 groups) PID temperature control separate segments, curve display and save the function display PV curve 0-8 days. Temperature can be set according to the linear operation (liters can set per hour up / cooling rate). | ||||
Standard configuration |
|
225L Temperature Humidity Chamer Features:
225 Liters Test Chamber Referring Picture:
Appearance Didplay:
No. | Name | Explanation |
1 | Observation window | For the customer to observe the status of the object which is being tested |
2 | Test area | Area for putting the test object |
3 | Controlling table | Interface for operating the machine |
4 | Right side of the machine | Build-in power device |
5 | Door handle | For fixing and connecting the door and machine |
Work principle:
1. The controlling system for adjusting temperature and humidity evenly controls SSR through PID. It makes the heating and moisture content equal the loss of heating and moisture
2. The dry and wet balls are used for measuring temperature signal
by comparing the A/D conversion input controller CPU and RAN, and
input to I/O board. The I/O board gives commands and instructions
to operate air circulation system and cooling and freezing system,
meanwhile, PID controls SSR, or heats or humidifies SSR for
operation in order to make the hot and wet content deliver to the
test box evenly through air circulation system, and finally achieve
the purpose of adjusting the temperature and humidity evenly