Add to Cart
Instruction:
PM-3000 Intelligent Pressure Gauges have full electronic structure
and use battery-powered, so they are very
easy for on-site installation. Specifically, front-end
high-precision piezoresistive pressure sensor is used and the
output
signal is processed by the high-precision, low temperature drift
amplifier, then sent into the high-precision A/D
converter to be converted into digital signal, after computing and
processing the actual pressure value is showed on
the LCD display.
Features:
The housing is water proof , widely applied for outdoor application
Accuracy can reach up to 0.25%
4 1/2 local display
Using slot for fixed , needn't srews at all , style designed by
America company
Applications:
Hydraulic/Pneumatic Control Systems
Energy and Water Management
Automation equipment
Process control
Chemical
Drawing:
Digital Pressure Gauge with Battery-Powered
Operating Manual
1.Summary
PM-3000 Digital Pressure Gauge Series are designed for pressure
applications that involve measurement of any media compatible with
316 stainless steel. Designed for battery operation, this digital
gauge has an automatic shutoff feature allowing the user to walk
away after taking a reading .Water-proof housing adopted for harsh
environment is our big advantage.
2.Technical Parameters
Wetted parts | SS304/ss316 | Operable Temp. | -20~85 |
Housing | Plastic (water proof) | Storage Temp | -20℃…125℃ |
Power supply | 9V battery | Long term stability | ±0.25%FS(annual) |
Media | Be compatible with 316L | Zero error(%FS) | ±0.5% |
|
| Full range error(%FS) | ±0.5% |
Thermal Zero drift(%FS) | ±1.0% | ||
Pressure port | G1/2,G1/4 optional | Thermal Span drift(%FS) | ±1.0% |
Life cycle | 9 millions | ||
Accuracy | 0.5%FS(typical) ,0.25%FS | Response | 5ms |
| 0…600 | Insulation | Min1MΩ @250VDC&25 |
Shock | 100g,11msec | ||
Over pressure | 3X Full Range | Noise | < 2m VRMS |
Burst pressure | 4X Full Range or 20kpsi which is less | Pressure reference | Absolute /Vent Gauge |
Life | >107Full Range Pressure | ||
Display | 4 1/2 digit LCD | Weight | 200g Max |
3.Parameters table
Name | Item | Setting range |
FL | LowerRange | -1999-19998 |
FH | UpperRange | -1998-19999 |
FLC | Zero Correcting Valve | -1999-1999 |
FHC | Full Scale Correcting Valve | 0.500-1.500 |
UNIT | Unit display optional | 0-15 |
DP | Decimal Point | 0-3 |
FT | Digital Filter Time Constant | 1-250 |
SPS | Sampling speed | 1-6 |
YPP | Peak Holding | 0-1 |
RUN | Automatic Power-off Setting | 0-1 |
TIMR | Automatic delay power-off time setting | 0-99.6H |
4. Operations
Name | Remarks | ||
Display Window | Display Window of measured Value | Indication of Measured Values. | |
| The S Key | In Measuring Mode, hold down the key for more than 3 seconds to
enter the Set Up Mode. | |
The on/off Key | Switching on press the key more than 3 seconds, if off , repeat . | ||
The back/light Key | Press the key more than 3 seconds to open back light . | ||
5. Installation
PM-3000 can be directly mounted in the hydraulic pipe through the
pressure pipe joints (M20*1.5) other size fittings available when
ordering .In critical applications , such as severe vibration or
shock , pressure hose pipe connection can be micro-mechanical
decoupling.
Note: The range less than 100Kpa must be installed vertically.
6. Battery Replacement
Remove back cover in the direction of arrow of the unit carefully
remove battery from the holders by lifting up the positive end of
the battery (opposite the spring) .Take care not to bend or distort
the battery retention springs. Do NOT discard the old battery into
fire, any other sources of extreme heat ,or in any other hazardous
manner. Always replace both batteries at the same time with high
quality alkaline laminated batteries. Observe the polarity of the
batteries when replace them. The negative (flat) end of batteries
should be inserted first and should face the spring in the battery
holder. Replace the back cover, including the rubber sealing
gasket.