Add to Cart
Intelligent Automatic Touch-Screen Control Oxygen Index Tester
I. Intelligent Full-automatic Touch Screen Control Oxygen Index
Tester Product Advantages:
1. Full-color touch screen control, simply set the oxygen
concentration on the touch screen, the program will automatically
adjust to the oxygen concentration balance and emit a squeaking
sound prompt, eliminating the need to manually adjust the oxygen
concentration;
2. The use of stepping proportional valves has greatly improved the
flow control accuracy. Closed-loop control is used. During the
test, the oxygen concentration drift program is automatically
adjusted back to the target value, avoiding the disadvantages of
the traditional oxygen index analyzer not adjusting the oxygen
concentration in the test. Improve the test accuracy.
II.Technical parameters
Control Method | Touch Screen |
Oxygen concentration range | 0~100% |
Digital resolution | ±0.1% |
Measurement accuracy | 0.1 level |
Flow adjustment range | 0-20L/min (0-1200L/h) |
Quartz glass tube | 75cm (can be customized according to user requirements) |
Gas velocity in the combustion tube | 40mm±2mm/s |
Pressure gauge accuracy | 0.01MPa |
Input pressure | 0.25-0.3MPa |
Working pressure | Oxygen: 0.15-0.20Mpa Nitrogen:0.15-0.20Mpa |
Specimen clamps can be used for soft and hard plastics, various types of building materials, textiles, fire doors, etc. | |
Propane (butane) ignition system, the ignition nozzle is made of a metal tube, the end of the inner diameter of Ф2±1mm nozzle, can be freely bent. Can be inserted into the combustion cylinder to ignite the specimen, flame length: 16±4mm, size 5mm-60mm can be freely adjusted. | |
Gas | Industrial nitrogen, oxygen, purity >99% |
Test benches | Bench not less than 1 metre long by 0.75 metre wide |
Ventilation requirements | Need to put the equipment into the fume hood or do the fume collection into the fume treatment and purification system. |
Test requirements | Dual gauge pressure reducing valve for cylinder (capable of regulating 0.2 MPa) |
Power | Maximum 150W |
Dimension | 650*400*830mm |
Weight | 30kg |
ASTM D 2863, ISO 4589-2, ISO 4589-3, NES 714, NES 715
Tips: Oxygen index tester requires not less than 98% of each
industrial-grade oxygen/nitrogen bottle as the gas source. Because
the above gas is a high-risk transport product, it cannot be
provided as an oxygen index analyzer fitting, and it can only be
used locally by users. Gas station purchase. (To ensure the purity
of the gas, please purchase it at the local regular gas station)
What is oxygen Index test?
Measuring the Minimum Oxygen Concentration to Support Candle-Like Combustion of Plastics (Oxygen Index) ASTM D2863. Scope: ASTM D2863 is a method to determine the minimum concentration of oxygen in an oxygen/nitrogen mixture that will support a flaming burn in a plastic specimen.