LK-F13L LED Light EMS / Woodpecker Compatiable Dental Ultrasonic
Scaler
Description:
This product is used for teeth cleaning and also an important
device for teeth disease prevention and treatment. The ultrasonic
scaler is composed of main unit, handpiece, cable, water pipe, tip,
torque wrench, foot switch and power supply.
The ultrasonic tooth scaler Technical Datas
Power source | 220V/110V |
Main unit input | 24V 50Hz ± 1Hz 1.3A |
Output power | 3~20W |
Frequency | 28±3KHz |
Half excursion force | 0.1N-2N |
Weight | 0.75kg |
Handpiece Model | HP-5L |
Water pressure: | 0.1bar~5bar(0.01MPa~0.5MPa) |
Compatible Brand | EMS |
Feature:
Detachable handpiece, with LED light
6 Colors Option: Red, Yellow, Orange, Blue,Green, Purple
Scaling and Perio Function included
Compatible with EMS and UDS
Automatic frequency tracking
Accurate feedback control and constant power output, ensure a more
comforatble feeling to patients.
Packing Information:
Packing: 5pcs/carton
Carton Size: 60*40*33/cm
FAQ:
1 The ultrasonic scaler worked fine before. But recently, power
decreased obviously. What is the reason?
First, check if ther is water or stains in the connection between
handpiece and cable. The water or stains will cause electricity
leakage, generate heat, decrease tip vibration, even stop water
atomizing properly. So, it is very important to make sure cable and
scaler dry before usage.
Solution: Dry handpiece and cable connector. Moist the two O rings
of the connector with a drop of clean water, then connect. Power
should be recovered mostly.
2 Handpiece cable connector is quite clean. But the power is still
very weak. What is the reason?
First, check if the tip is well screwed into the handpiece
If the tip is not connecting well, vibration cannot be passed to
it. Even you switch the power to max, tip is still very weak.
Atomizing is bad. Heat will also be generated because of rub from
handpiece and tip.
Solution: Screw the tip tightly. Before you screw the tip, you need
to unscrew it and clean the dirt around screws with alcohol cotton.
The dirt will also make tip loose. After that, screw the tip
tightly. Power will be recovered.
3 Why does the scaler tip become loose?
The tip is vibrating during working. It will be loose after a long
time vibrating. In fact, being loose is normal phenomenon.
4 Why my scaler tips always loosen? Sometimes it can’t even treat
one patient!
The tip is not screwed tightly. Generally there are two reasons:
One: there is dirt between the screw thread of tips and handpiece.
Two: Tip is not screwed tightly enough. You should screw it more
with torque wrench.
5 How do I know I use proper power to screw it?
Torque wrench should be used to screw the tips.
6 While handpiece connector is clean, and scaler tip is screwed
tightly,
however, tip vibration is still very weak. And water spray is weak,
too. Why?
A. Scaler tip might be cracked. Generally, new scaler tip has good
performance. So it works fine even if you switch the power into low
stage. If you switch the power to higher stage when you use new
tip, tip get worn very fast. So please switch to low power stage
when you use a new tip. As the tip get worn slowly, just increase
power accordingly. In this way, the tip can be used longer while
operators will not feel tip vibration get weaker.
B. Scaler tip is worn too much. As scaler tip gets worn gradually,
power stage has to be increased accordingly. After the power stage
is increased to max, the tip will get worn further, and there is no
way to compensate the vibration further. Vibration will be weaker
and weaker. The final and easy solution is to change a new tip.
7 Both vibration and water spray are good, but you can feel scaler
tip is heat when you touch it. Is this a problem?
No, it is not a problem. As the scaler is driven by ultrasound, the
tip vibrates, which generate hole effect. Hole effect from
vibration will clean stains and calculus fast. It is noticed that
the calculus is not scaled by scaler, but moved by vibration.
Nerves of gingival, lips, fingers etc can feel heat or hurt during
vibration. That is the reason. Generally, it will not burn people.
Heat from finger proves that scaler tip’s power, not problem.
8 The machine works fine. But scaler handpiece connector is full of
rust. Will this bring trouble to machine itself?
Yes. This connector is to transfer water and electricity. Little
rust will not make any influence. Too much rust will not only
affect circulation of electricity, but also cause electricity
leakage, even burn the whole circuit.
9 How to avoid the rusting of the scaler handpiece connector?
Solution:
A: Make the scaler handpiece connector dry with 3 way syringe. Then
put one drop of water into the rubber O ring.
B: Please make sure waterproof cover buckled to prevent the
disinfectant corrode metal parts when the handpiece are
disinfecting.
10 The scaling tip doesn’t vibrate but there is water flowing out
when stepping on the switch. And sometimes the tips will not
vibrate until you touch it.
Solution: Please use compatible scaler tips.