IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant Battery

Brand Name:LIB
Certification:CE,ROHS
Model Number:TK10-220
Minimum Order Quantity:1 Set
Delivery Time:5-14 Days
Payment Terms:L/C, D/A, D/P, T/T, Western Union, MoneyGram,Paypal
Contact Now

Add to Cart

Verified Supplier
Address: No.1 Zhangba First Street, Xian High-Tech Area, Shaanxi, P.R. China 710065
Supplier`s last login times: within 42 hours
Product Details Company Profile
Product Details

Description

IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber are suited for testing battery reliability under extreme temperature changes.

Battery need to be able to withstand drastic changes in temperature.

The thermal shock and temperature cycling tests called out in the IEC, UL, and SAE standards


Technical Specification

NameIEC 62660-2 130℃ Battery Temperature Cycling Test Chamber
ModelTK10-220
Internal Dimension (mm)500*600*750
Overall Dimension (mm)960*1150*1860
Temperature Range (℃)-70℃ ~+150 ℃
Cooling Rate10℃/min (15℃/min)
Heating rate10℃/min (15℃/min)
Temperature Fluctuations± 0.5 ℃
Temperature Deviation± 2.0 ℃
Interior Volume220 liters
ControllerProgrammable color LCD touch screen controller, Multi-language interface, Ethernet , USB
Cooling SystemMechanical compression refrigeration system
Exterior MaterialSteel Plate with protective coating
Interior MaterialSUS304 stainless steel
Power Supply380V
Maximum Noise65 dBA

Thermal Test

Battery Temperature Cycling Test Chamber,This test is performed to characterize cell responses to high-temperature environment.

a) Adjust the SOC of the cell to 100% for BEV application, and to 80% for HEV application.


b) The cell, stabilized at room temperature, shall be placed in a gravity or circulating air convection oven. The oven temperature shall be raised at a rate of 5K/min to a temperature of 130℃±2K. The cell shall remain at this temperature for 30 min before the test is discontinued.


Note:

If necessary, to prevent deformation, the cell may be maintained during the test in a manner that does not violate the test purpose. The manner to prevent deformation should be representative of cells inside battery systems and battery packs.


IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber


Test Standard

1. IEC 62660-2:The oven temperature shall be raised at a rate of 5K/min to a temperature of 130℃±2K.

2. IEC 62133 - Temp cycling: 75 to 20 to -20°C (30 min. transitions)

3. UL 2054 - Temp cycling: 70 to -40 °C (30 min. transitions)

4. UL 1642 - Temp cycling: 70 to 20 to -40 °C (30 min. transitions)

5. SAE J2464 - Thermal shock test: 70 to -40°C (15 min. transitions)


The LIB Battery Temperature Cycling Test Chamber meet IEC 62660-2 standard for test battery