Setting Up the unit:
Verify that the Power On/Off Switch is in the Off position and that
no accessories are connected to the unit.
Connect a hospital grade power cable to the AC power cable
receptacle on the back of the unit, then to a properly grounded
wall outlet.
Turn the unit on by switching the power switch to the On position.
Connect a two-pedal footswitch to the appropriate receptacle on the
rear of the unit. Use only MECHAN’s footswitches. Although other
types of footswitches may fit, they may not be compatible.
Connect a treatment handle and electrode at this time
Fire Explosion Hazard:
The following substances will contribute to increased fire and
explosion hazards in the operating room:
• Flammable substances (such as alcohol based skin prepping agents
and tinctures)
• Naturally occurring flammable gases which may accumulate in body
cavities such as the bowel
• Oxygen enriched atmospheres
• Oxidizing agents (such as nitrous oxide [N2O] atmospheres)
The sparking and heating associated with electric surgery can
provide an ignition source. Observe fire precautions at all times.
When using electric surgery in the same room with any of these
substances or gases, prevent their accumulation or pooling under
surgical drapes, or within the area where electric surgery is
performed.
>