KFDO900 Do Meter Stainless Steel Sensor Water Quality For
Aquaculture Industrial
Product Introduction
dissolved oxygen measuring instrument is based on fluorescence
quenching principle.The fluorescent material at the front of the
sensor is a special platinum porphyrin compound with a polyester
foil that allows gas molecules to pass through. The surface is
coated with a black light-blocking material to avoid interference
by sunlight and other fluorescent substances in the water.
When blue light hits the fluorescent material, the fluorescent
material is excited and emits red light. Since oxygen molecules can
carry away energy (quenching effect), the time and intensity of the
excited red light are inversely proportional to the concentration
of oxygen molecules.By measuring the phase difference between the
excitation red light and the reference light and comparing with the
internal calibration value, the concentration of oxygen molecules
can be calculated.After linearization and temperature compensation,
the final value is output.
The advantage of fluorescence dissolved oxygen
In general, the fluorescence dissolved oxygen instrument can
measure the oxygen content of water without flow rate, which is
more accurate without the need for magnetic stirring
machine.Although the first purchase cost is high, the electrode is
very easy to use. It is usually replaced every 2-3 years, unlike
the polarographic method, which requires electrode replacement, and
the cost is very expensive.
Performance characteristics:
● Innovative fluorescence principle, no need to replace the
diaphragm and electrolyte, basic maintenance free;
● No need to replace the fluorescent cap, lower use cost;
● No oxygen consumption, no flow rate and agitation requirements;
● will not be "toxic" because of sulfide, to overcome environmental
interference;
● Free from cross interference of H2S, PH, CO2, NH3, SO4, CL-, CL2,
etc.
● Higher resolution and measurement accuracy;
● The service life of the sensor is up to 5 years;
Apply:
Surface water, industrial sewage and municipal sewage
Q&A
What is zero drift?
It refers to the size of the indicating value of the automatic
analyzer changing within a certain period of time when the zero
correction liquid specified in this technical requirement is used
as the sample for continuous testing.
What is Range drift?
refers to the change of the indicating value of the automatic
analyzer in a certain period of time when the range correction
liquid specified in this technical requirement is used as the
sample for continuous testing.
What is Mean fail-free?
continuous running time refers to the ratio of the total running
time (h) of the automatic analyzer to the number of failures
(times) during the test period, which is shown in "MTBF". The unit
is h/time.
What is response time (T90)?
electrodes from range correction fluid to zero correction fluid,
the determination of displayed value range correction liquid
concentration 90%.
Technical Parameters of Fluorescence Dissolved Oxygen Analyzer
Class | Performance | Parameter |
Transmitter | Size | 230×185×120mm |
Weight | 1.5kg | |
Storage Temperature | -40-60℃ | |
Temperature | -20-60℃ Avoid direct sunlight | |
Relative humidity | 5%~85% (no condensation) | |
Supply | 220VAC±10%,50Hz | |
Analog output | 4~20mA output or 0~20mA output (dissolved oxygen value or saturation), loop load maximum 750 ohms;Optional temperature 4~20mA output | |
The relay outputs | 3 relays with A capacity of 220 VAC/ 2 A. The outputs can be set for alarm or timing functions | |
Protective grade | polycarbonate enclosure, IP65 | |
Field display | LCD display with backlit | |
Digital interface | RS232 or RS485 interface, Modbus protocol can be supported. | |
Sensor | Measurement range | dissolved oxygen: 0.00~20.00mg/ L or 0.00~20.00 PPM air saturation: 0~200% |
Resolution | 0.01mg/L | |
The measurement accuracy | ±1%FS or ±0.15mg/L with the largest value | |
Repeatability | ≤0.2mg/L | |
Response time | less than 60 seconds | |
Temperature sensor | Working range: 0~60°C;Resolution: 0.01°C Accuracy: ±0.5°C;Response time: less than 10 seconds | |
Storage temperature | -20°C ~60°C | |
Ambient temperature | -20°C ~60°C | |
Medium temperature | 0℃~50℃ | |
Maximum mounting depth | 10 m | |
Chemically adaptation | methanol and ethanol are safe to avoid other organic solvents | |
Calibratation | Anoxic water and air were used for calibration respectively.It has been calibrated before leaving the factory, and generally there is no need to calibrate. | |
Anti-interferense | It is not subject to cross interference of H2S, PH, CO2, NH3, SO4, CL-, CL2, CLO2, MeOH, EtOH, and various other ions. | |
Foil | Polyester embedded with platinum porphyrin fluorescence | |
Light-proof coating | teflon | |
Foil life | up to 5 years (10 second measurement interval) | |
Sensor shell size | 45 x 100 Φ | |
Weight | 1.2 kg | |
Cable standard | 10 meters | |
Sensor shell | stainless steel |
Ordering Information selection table
Ordering Information selection table | ||||
KFDO900 | Fluorescence Dissolved Oxygen Meter | |||
Power supply | ||||
DC24V | ||||
AC220V | ||||
Cable length | ||||
10m | ||||
15m | ||||
20m | ||||
30m | ||||
Signal output | ||||
Only 4-20mA | ||||
Only 485 | ||||
Also with 4-20mA , 485 signal output | ||||
Special custom signal | ||||
KFDO900- | AC220V | 10m | Only 485 | with AC220V adapter, cable between host and sensor,The cable length
is 10 meters, the sensor is equipped without temperature
compensation, and is equipped with only 485 communication signal
output. |