KCL501 Online Chlorine Ion Sensor Signal Output: RS-485 Bus Patented Chloride Ion Electrode Easy To Install: 3/4”NPT
1.Application Environment
The KCL501 integrated on-line chloride ion sensor uses a solid-membrane Ion selective electrode to measure free chloride ions in water quickly, easily, accurately and economically. The technical parameters, communication protocol and operation and maintenance of the chlorine ion sensor are described in detail in this manual.
2.Feature
3.Technical Specifications
Model number | KCL501 |
Range Range | 0 ~ 3500mg/L or 0 ~ 35000mg/L |
Resolution | 0.1 mg/L or 1 mg/L |
Precision | ± 5% |
Operating temperature | 0 ~ 40 °C |
Working pressure | < 0.1 mpa |
The pH range of the medium | 2 ~ 12ph |
Temperature compensation | Automatic temperature compensation (PT1000) |
Power supply | 12 ~ 24 VDC ± 10% |
Signal output | RS-485 bus, Modbus/RTU protocol |
Contact material | POM |
Installation mode | 3/4”NPT pipe thread, immersion mounting.” |
Cable length | 5M, other length can be customized |
Calibration mode | Two-point calibration |
Power consumption | < 0.3 W@12 V |
Protection level | IP 68 |
4.Dimensional Drawing
Note: Sensor Connector is M16-5 core waterproof connector
5.Installation and Electrical Connection
Installation
Note: The sensor can not be installed upside down or horizontally, at least tilt
more than 15 degrees angle installation.
Electrical Connection
The cable is a 4-core twisted pair shielded wire defined in line sequence:
Red wire -- power cord (12 ~ 24 VDC)
Black line (GND)
Blue Line -- 485 a
White line -- 485 b
Before electrification, the wiring sequence should be carefully checked to avoid
unnecessary losses caused by wiring errors.
Wiring instructions: Consider the cable long-term immersion in water (including seawater) or exposed to air, all wiring are required to do waterproof treatment, customer cable should have a certain anti-corrosion ability.
6.Maintenance
Remove the protective cap on the electrode head. Note: Do Not Touch the sensitive part of the electrode tip with your finger.
The electrode should be kept dry before use. The sensor of the electrode should be put into the protective bottle. Before testing, the electrodes should be immersed in running water for 24 hours with electric current. If stored overnight or longer, the electrode head should be cleaned with deionized water, dried, and then placed in the original packaging.
Check whether the terminal is dry, if there is any dirt, please wipe with absolute alcohol, blow dry after use. Long-term immersion in distilled water or protein solutions should be avoided and contact with silicone oils should be avoided. The electrode is used for a long time. If there is any measurement error, it must be corrected with the instrument.
When the electrode can not be calibrated and measured after being maintained in the above way, please replace the electrode.
Interference ion: Sulfur Ion, iodine ion, bromine ion, etc. .
Note: The sensor has been calibrated before the factory, if not out of measurement error should not be arbitrary calibration.
Zero calibration
Place the sensor in a vial containing a standard solution of zero point and wait 5 minutes for the value to stabilize to see if it is within the range of error, or if it is not, zero calibration is required. Reference Appendix to calibration instruction.
Slope calibration
Place the sensor in a vial containing the slope standard solution and wait 5 minutes until the values are stable to see if they are within the error range. If not, slope calibration is required. Reference Appendix to calibration instruction.
7.Quality and Service
Quality Assurance
Quality inspection departments have standardized inspection procedures, with advanced and perfect testing equipment and means, and in strict accordance with the rules of inspection, 72 -hour
aging of products to do experiments, stability experiments, do not let a unqualified product factory.
The consignee will return the lot of the product with 2% nonconformity rate, and all the cost will be borne by the supplier. The standard of consideration refers to the product description provided by the supplier.
Guarantee the quantity and speed of delivery.
Spare Parts
This product includes:
1 sensor
1 copy of instruction manual Certificate 1
According to the range, 100 mg/L and 1000 mg/L calibration solution were selected as 100ml or 1000 mg/L and 10000 mg/L standard solution as 100 ml respectively
After-sales Service Commitment
The company provides six months from the date of sale of this machine after-sales guarantee. But does not include the damage caused by improper use, if need repair or adjustment, please send back, but the freight is at your own expense, send back to make sure the packing is good to avoid damage in transit, our company will repair the damage free of charge.