Dual High Impedance KPH601 Online PH Sensor Differential Amplifier

KPH601 Online PH Sensor Signal Output: RS-485 (Modbus/RTU Protocol)

1.Calibration

Both the zero point, the point where the pH electrode delivers 0 mVpotential and the slope of the calibration line show manufacturing dependent tolerances and will change after exposure to the measuring solutions. Therefore the pH electrode has to be calibrated with accurately defined buffer solutions. In order to make an exact calibration, the zero point of the measuring chain has to be known. This is generally at pH 7. Two buffer solutions chosen according to the desired accuracy and measuring range should be used. The value of the first buffer solution should be close to zero mV (at pH 7). The second buffer solution should have a pH value within the measuring range. It should be noted that the two buffer solutions must have a difference in pH values of at least two pH units. As already mentioned, the pH value is temperature dependent. Therefore, it is important that the temperature curve of the buffer is known. Additionally, the buffer and the electrode must be at the same temperature, or you must wait for the temperature to reach equilibrium. The zero point calibration (pH 7, first buffer) always has to take place before the slope calibration (with a pH value close to the measuring value, second buffer) except when working with a microprocessor controlled system. If a very accurate measurement is required, it is advisable to repeat the zero point calibration after the slope calibration.


The following factors directly influence the accuracy of the calibration as well as the pH measurement:

• Buffer solutions

• Temperature measurement and temperature compensation

• Condition of the junction and the reference system (contamination, etc.)

• Working technique

The stability of the zero point and the slope depends on the composition of the measuring solution as well as on the temperature. It makes little sense to make general statements regarding the calibration frequency. Therefore, when working with unknown solutions it is advisable to repeat the calibration often in the beginning. When the calibration values are stable, the calibration frequency may be extended. In general the calibration frequency depends on the desired accuracy. To be able to calibrate during the process, an electrode in a retractable housing is very useful.

2.Technical Specifications

ModelKPH601
Measuring Range0~14pH
Resolution0.01pH
Accuracy±0.1pH
Working Temperature0~65℃
Working Pressure<0.2MPa
Temperature CompensationAuto temperature compensation (NTC)
Power Supply12~24VDC ±10%
Signal OutputRS-485(Modbus/RTU)
Wetted MaterialPOM
Mounting MethodImmersion mounting, 3/4 NPT thread
Cable Length5 meters, other lengths can be customized
CalibrationTwo-point calibration
Power Consumption<0.3W@12V
Protection GradeIP68

3.Dimensional Drawing

Note:The sensor connector is M16-5 core waterproof joint male head.

4.Installation

Note: The sensor should not be installed upside down or horizontally when installed, at least at an angle of 15 degrees or more.