Add to Cart
5 Red LED digital tubes for English&digit display With Special Anti-vibration Digital Filtering Algorithm For Precise Weighing
No. | Pin | Description | |
LOADCELL | Loadcell Port | ||
Connection Mode | 6-Wire | 4-Wire | |
Internal Dip Switch S4 | S4→【6-WIRE】 | S4→【4-WIRE】 | |
1 | SIG- | Weighing Signal [mV] Input -. | Weighing Signal [mV] Input -. |
2 | SIG+ | Weighing Signal [mV] Input +. | Weighing Signal [mV] Input +. |
3 | EXC- | Excitation Voltage -. | Excitation Voltage -. |
4 | SEN- | Feedback Voltage -. | |
5 | EXC+ | Excitation Voltage + [DC9V]. | Excitation Voltage + [DC9V]. |
6 | SEN+ | Feedback Voltage +. | |
7 | SHD | Shield Ground. | Shield Ground. |
POWER | DC24V[±20%] Power Input Port | ||
1 | DC- | DC Input -. | |
2 | DC+ | DC Input +. | |
The metal shell should be grounded directly. For separating the controller from the interference of the driving devices, the DC24V power supply of the controller should not be shared by the DO. |
Pictures
Technical Specifications
Power Supply: DC24V±20%, Max. 5W.
Loadcell Excitation Voltage/Current: DC5V/120mA.
8 Loadcells[350Ω] connectable.
Transmitter Excitation Voltage/Current: DC12V/100mA.
Optional weighing input signal mV, V and mA.
【mV】Signal Input Range: 0~19.5mV.
【V】Signal Input Range: 0~2.5V, 0~5V, 0~10V.
【mA】Signal Input Range: 0~20mA.
2 Normally open/closed Relays [DO]: AC250V/DC24V, 1A.
1 Analog Signal Output [AO]: 0~20mA, 0.05%FS.
COM1: RS232.
COM2: RS485.
Connect Host IPC[Modbus] & Remote Display.
Outline Size [W×H×D]: 107 × 60 × 100 mm.
Panel Cut-out Size [W×H]: 94 × 47 mm.
Operating Temperature: -25℃~+45℃.
Protection Level of Front Panel: IP65.
Accuracy Grade: III.
Verification Accuracy: 0.03%.
Static Weighing / Force Measuring Accuracy: 0.2%~0.5%.
Name | Description | Range | Note |
X | Channel X Real-time Force Value | -5000~+5000 | Refer to Parameters [202]~[207]. |
Y | Channel Y Real-time Force Value | -5000~+5000 | |
Z | Channel Z Real-time Force Value | -5000~+5000 | |
[F] | Real-time Resultant Force or Resultant Force Peak Value [Force Value Unit]. | 3-D Algorithm: F2 = X2 + Y2 + Z2 |
LED light | Description |
[RUN] | OFF: Real-time Force Value tracking display state. |
[AL1] | Real-time Resultant Force or Resultant Force Peak Value Upper Limit
Alarm. |
[AL2] | Real-time Resultant Force or Resultant Force Peak Value Lower Limit
Alarm. |