Add to Cart
ISO 6502 Viscometer Rubber Viscosity Measuring Instruments
Use:
This machine is designed by GB/T16584<Rubber——use no-rotor rheometer to test vulcanization characteristic,can meet ISO6502 standard,and can test the data of Italy standards about T10,T30,T50,T60,T90.This machine can be use test unvulcanized rubber characteristic,and test out the suitable vulcanizing time. Used imported intelligent digital temperature controller,easy adjustment,wide temperature controlling range,high accuracy,good stability and reproducibility.Use computer to control the machine and store the test data and print the data.Test curve can be compared and amplification in software.
Main functions:
Using computer control, after setting the parameters on the computer, directly control the test parameters of the vulcanizer. Real-time display of vulcanization curve and temperature curve, storage of test results, can adjust the comparison of different test results and display them in different colors.
It adopts single-chip microcomputer control, which includes: host, temperature measurement, temperature control, data acquisition and processing, sensors and electrical interlocking components. Among them, the temperature measurement and control circuit is composed of a temperature controller, platinum resistance, and a heater, which can automatically track changes in the power grid and ambient temperature, and automatically correct PID parameters to achieve rapid and accurate temperature control. The data acquisition system and the electromechanical chain complete the automatic detection of the torque signal of the rubber vulcanization process, and automatically display the temperature and set value in real time. After vulcanization, automatic processing, automatic calculation, printing of vulcanization curve and process parameters. It displays vulcanization time, vulcanization torch, and has various audible prompts.
Features:
1. This instrument is really confined mold cavity, and the United
States Alpha (formerly Monsanto) has done. Repeatability, and test
data comparable with the Alpha. At the international leading
position in the same industry.
2. The instrument development platform based on large databases,
temperature control devices using the software to directly control
and collection and processing. Instrument to overcome the general
curing temperature of the shortcomings of using temperature
controller (accuracy poor). The technical leadership of
international new trend.
3. This instrument has statistics, analysis, storage and comparison
functions. Humanized design, easy to operate
4. Using imported high-precision sensors
How it works:
The rubber sample into the mold cavity is almost entirely enclosed
and maintained at test temperature, the mold cavity are of two
parts, of which the lower part with a small linear reciprocating
movement (swing oscillation), oscillatory shear specimen produced
strain determination is the reaction torque of the mold cavity
(force), this torque (force) depends on the shear modulus of
rubber.
Curing test specimens after the start of the shear modulus
increases, the computer machine real-time display and record the
torque (force), when the torque (force) rose to a stable value or
the maximum as well as return to the state, they get a torque (
force) and time curve, that is, curing curve (Figure 1), shape of
the curve and test the temperature and plastic material
characteristics.
Download Brief:
ZL-3001 No Rotor Rheometer.pdf
The main technical parameters:
Model | Without Rotor Rheometer Rubber Testing Machine |
Temperature tester | Room temperature to 200°C |
Temperature accuracy | ≤ ±0.3°C |
Temperature resolution | 0.1°C |
Torque range | 0-5N.M,0-10N.M,0-20N.M |
Torque resolution: | 0.001NM |
Motor rotating speed | 100 times / min |
Rotor range | Adopt 4 setting: 25, 50, 100, 200lbs per inch |
Die cavity frequency | 1.7Hz |
Sway angle | ±0.5°, ±1°, ±3° |
Air compression | 0.5Mpa--0.65MPa |
Ambient temperature | 0-35°C |
Test result | Printing:date,time,temperature,vulcanization curve,temperature curve,ML,MH,ts1,ts2,t10,t50,Vc1,Vc2. |
Dimension | 131*56*62cm |
Power | Single phase, 220V±10%, 50Hz |
Display | Windows 7 + software |