1. When used in a Class 100 laimar flow hood, the ultraviolet
germicidal lamp should be turned on 50 minutes in advance to deal
with the microorganisms accumulated in the clean workbench. After
30 minutes, the germicidal lamp should be turned off and the blower
should be started.
2. Do not store unnecessary items on the surface of the 100-level
clean laimar flow hood to keep the net airflow in the work area
undisturbed.
3. Newly installed or unused class 100 clean benches must be
carefully cleaned with ultra-clean vacuum cleaners or
non-fiber-producing tools for the clean benches and surrounding
environment before being used, and then UV sterilization Carry out
sterilization treatment.
4. Always pay attention to the indication of the differential
pressure gauge above the working area. When the pointer enters the
red zone from green, it indicates that the dust holding capacity of
the high-efficiency air filter is becoming saturated. The fan
voltage should be increased, and the wind speed in the working area
must be measured with an anemometer. When the average wind speed is
less than 0.3m/s, then The filter has expired and should be
replaced.
5. Unnecessary items should not be stored in the clean workbench to
keep the net air flow pattern undisturbed.
6. After installing the Class 100 clean bench, please check again
whether all parts are normal or not. Before starting the work,
please clean it thoroughly with alcohol. Check whether the power
plug is reliably inserted into the power socket. Turn on the power
switch, the controller power indicator light is on, press the
increase/decrease button of the fan to start the operation of the
fan, please do not put items in the suction port or above the
suction port, so as not to cause damage to the product due to the
reduced air intake performance.
When the resistance of the high-efficiency filter is greater than
200Pa or less than 0.3m/sec, the high-efficiency filter should be
replaced.
7. Always pay attention to the indication of the differential
pressure gauge above the working area. When the pointer enters the
red zone from green, it indicates that the dust holding capacity of
the high-efficiency air filter is becoming saturated. The fan
voltage should be increased, and the wind speed in the work area
must be measured with an anemometer. When the average wind speed is
<0.
At 3m/s, the filter has failed and should be replaced.
8. Regularly use a hot bulb electric anemometer to measure the
average wind speed in the work area. If it is found that the
technical indicators are not met, the voltage of the fan unit motor
can be adjusted to make the class 100 clean workbench in a better
working condition.
9. The HT533A air volume controller is used to explain that the
HT533A air volume controller is one of the HT533 series of air
volume controllers. It is a special modulator designed and
developed to adjust the air volume for the clean bench.
There are eight lines of electrical appliances, adjustable voltage
memory. Touch keyboard operation and other functions.
10. The class 100 clean bench can be operated by one person or by
two persons.
W*D*H | Outside:1040*740*2000 Inside:950*645*800 | Outside:1340*740*2000 Inside:1250*645*800 | Outside:1640*740*2000 Inside:1550*645*800 |
Suitable number | 1 | 1-2 | 2-3 |
Air quantity(m3/h) | 905 | 1205 | 1510 |
Air speed(m/s) | 0.45m/s±20% | ||
Cleanliness | Class 100(US Federal standard 209E) | ||
HEPA Eff | 99.99%@0.3um | ||
HEPA size | 915*610*70-1pcs | 1220*610*70-1pcs | 915*610*70-1pcs 610*610*70-1pcs |
Working surface material | Sanding stainless steel SUS304 | ||
Noise | <56dB | <56dB | <58dB |
The cleanbench is 0.2m in front and 1.2m above the ground | |||
Power source | 220v 155w | 220v 175w | 220v 275w |