

Add to Cart
DJY-4 ECG Electrode Electrical Performance Tester
Overall dimensions: 27.5cmx22cmx10cm.
Total machine weight: 2.5Kg.
Fuse: F2A 250V.
Technical parameters:
Test items | Incentive conditions | scope | Accuracy | Testing time |
Impedance | Sinusoidal alternating current | 0-20KΩ | Less than or equal to 1KΩ 2%±10Ω | >60s |
Disorder | DC input impedance | ±1200mV | Less than or equal to 100mV 1%±1mV | 60-90s |
Noise | RC Bandpass | 0-2000μV(PP) | Less than or equal to 150μV ±7.5μV | 60-360s |
Flow resistance | DC Current | ±1200mV | Less than or equal to 100mV 1%±1mV | More than 8 hours |
Defibrillation | Capacitor 10μF±5% | ±1200mV | Less than or equal to 100mV 1%±1mV | 5, 15, 25, 35s |
Environmental conditions
a) Power supply voltage: AC 220V±10%, 50Hz±1Hz, with good
grounding.
b) Ambient temperature: 23℃±5℃.
c) Relative humidity: 40%±10%.
d) Atmospheric pressure: 70.0~106.0Kpa.
Standard configuration: 1 main unit, 1 external power supply, 1
instruction manual, 1 grounding wire, 1 pair of 200nA connections,
15 electrode pair series wires, 1 portable case
Standard:
YY/T 0196-2005 Disposable ECG Electrodes
ANSI/AAMI EC12:2000(R2005)
Application Scope:
Used for testing the electrical properties of ECG electrodes ,
suitable for disposable body surface ECG electrodes , including
disposable ECG electrodes with pre-installed conductive glue or
self-adhesive ECG electrodes with pressure-sensitive conductive
glue and disposable ECG electrodes with conductive glue placed
according to the requirements before use .
ANSI/AAMI EC12:2000(R2005) ,YY/T O196-2005 "Disposable ECG
Electrodes " stipulates the requirements for batch inspection and
periodic inspection, that is, the grouping of electrode pairs and
the test sequence of each item. The requirements for the above test
items are as follows:
Test items | Impedance | Disorder | Noise | Flow resistance | Defibrillation |
Industry Standard Requirements | < 3KΩ | < 100mV | < 150μV | < 100mV | < 100mV |
AC Impedance ("Impedance" for short)
A comprehensive measurement indicator that represents the impedance
and capacitive reactance of an electrode pair.
During the test, the instrument applies a 10Hz sinusoidal AC
current with a peak-to-peak value not exceeding 100μA to the
electrode pair through two electrode test cables. In this case, the
AC impedance of the electrode pair should not exceed 3KΩ.
DC offset voltage (offset for short)
The voltage developed between two disposable ECG electrodes due to
the difference in their half-cell potentials.
After forming the electrode pair, after a 1 minute stabilization
period, the DC offset voltage of the electrode pair measured
through the electrode test cable should be no more than 100mV.
Internal Noise (“Noise” for short)
The voltage signal within a certain frequency band is presented by
the change of the offset voltage of the electrode pair.
After forming the electrode pair, after a 1 minute stabilization
period, within the following 5 minutes, the peak-to-peak value of
the internal noise generated by the electrode pair in the frequency
band of 0.15-100 Hz (first-order frequency response) measured by
the electrode test cable should be no more than 150μV.
Bias current tolerance (abbreviated as "current tolerance")
The ability of an electrode pair to maintain a stable DC offset
voltage under the long-term action of a given small DC current.
The electrode pair is connected to the 200nA terminal of the
instrument, and then a 200nA DC current is continuously applied to
the electrode pair for more than 8 hours. During the entire period,
the voltage change across the electrode pair measured by the
electrode test cable should not be greater than 100mV.
Simulated Defibrillation Recovery Performance (abbreviated as
"Defib")
The ability of the electrode pair to restore ECG detection after
being stimulated by defibrillation current.
During the test, the 10μF capacitor in the instrument is charged to
DC200V, and the electrode pair and the 100Ω resistor in the
instrument are discharged in series through the electrode test
cable. The voltage value on the electrode pair should be no more
than 100mV in the 5th second after the capacitor begins to
discharge; within the next 30 seconds, the average rate of change
of the voltage every 10 seconds should be no more than ±1mV/s. It
is also required that after four times of the above experiment, the
10Hz AC impedance of the electrode pair should be no more than 3KΩ.