Medical Laboratory Equipment Class II biological safety cabinet


Application Range:


Biological safety cabinet is indispensable equipment for modern microorganism lab, especially for those occasions that the operators need to take protective measure, such as medicine, pharmacy, microbiology, biomedicine, gene recombination, animal experiment and biological products. It can provide a sterile clean and safe environment.


Structural Characteristics:


1.BHC-ⅡA2 constant airemissions by 30%,BHC-ⅡB2 constant airemissions by 100%, negative pressure vertical laminar flow.

2.Movingup / down glass door, flexiblepositioned, and can completely shut down for sterilization, alarm prompt.

3.High efficiency filters to filter particles, dust and microorganism in the air,two layers of filter for more than 99.9999% of the HEPA filter membrane

4.Usinghigh quality 304 stainless steel,can prevent the corrosion and erosion of disinfectants.

5.Equipped with an alert function forhigh efficiency filter failure

6.The alarm forair leakagein workspace.


Technical Parameters:


Model

Item

BHC-1000II A2

BHC-1000IIB2

BHC-1300II A2

BHC-1300IIB2

BHC-1600II A2

BHC-1600IIB2

Clean levelISO 5(Class 100)/ISO 4(Class 10)
HEPA Filter Efficiency@0.3~0.12µm≥99.999% ~ 99.995%
Downflow Air Velocity0.25m/s~0.5m/s
Inflow Air Velocity≥0.5m/s
Noise≤60dB(A)
Vibration Half Peak≤5µm
Electric PowerAC,Single-phase 220V/50Hz(AC,220V,1¢,50Hz)
Laminar Flow ProtectionOperation Protection(Repeat 3 times)a.Total colony in impaction sampler≤10CFU/time
b.Total colony in impaction sampler≤5CFU/time
Sample Protection(Repeat 3 times)Total colony in culture dish≤5CFU/time
Protection Against Cross-Contamination(Repeat 3 times)Total colony in culture dish≤2CFU/time
Maximum Power≤1500KVA(with spare socket)
Weight(kg)150200250
Operation Room Dimension(mm)1000×570×6351300×570×6351600×570×635
Overall Dimension(mm)1200×855×23501500×855×23501800×795×2350
Fluorescent Lighting Dimension & Quantity30W×②40W×②40W×②
UV Lighting Dimension & Quantity20W×①30W×①40W×①
Luminance≥650Lx
Exhaust Pipe’s DiameterΦ200Φ250
Exhaust DirectionEjection