First Aid Equipment Real Time Waveform Display Ventilator With 10.4 Inch LCD Display

Brand Name:MeCan
Certification:ISO
Model Number:PA-700 B
Minimum Order Quantity:1 set
Delivery Time:3-15 working dates
Payment Terms:L/C, D/A, D/P, T/T, Western Union, MoneyGram
Contact Now

Add to Cart

Active Member
Location: Guangzhou Guangdong China
Address: RM718, EASTERN BUILDING, XIAOBEI, YUEXIU, GUANGZHOU, CHINA.
Supplier`s last login times: within 36 hours
Product Details Company Profile
Product Details

Frist Aid Equipment CE Marked Real Time Waveform Display Ventilator


Model : PA-700 B


The PA-700B ventilator is gas driven by medical compressed oxygen and compressed air. In the inspiratory phase, two streams of compressed gas (compressed oxygen and compressed air) flow into a high performance air-oxygen mixer to form a mix of oxygen and air with a certain pressure.


Component Details of Ventilator


Display: 10.4''LCD Display Following Parameters & Real time waveform display

Operating mode and respiratory rate;


——Ventilator control frequency set by medical staff;


——Tidal volume of each breath in the patient;


——Occurrence of spontaneous breath in the patient and spontaneous breathing frequency;


——Expiratory phase and inspiratory phase and actual respiratory rate;


Ventilator Construction:


--Environmental Protection Material construction, hardly to be broken.

-- Trolly: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions


TECHNIQUES SPECIFICATIONS
Physical specifications
Display10.4'' LCD Display
Mode:pneumatically powered and electrically controlled system
Trolley:Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Technical Specifications:
Gas requirement:Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa
safety valve<12.5 kPa
Respiratory Rate1 ~ 99 bpm
Oxygen Concentration21% ~ 100%
Maximum Minute ventilation≥18L / min
modes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH, SPONT
Ventilator compliance≤30 Ml/kPa
PEEP:0 ~ 1.0 kPa
Ptr:-10 ~ 10 hPa
IMV frequency:1~20 bpm
I/E Ratio:4:1~1:4, Has the inverse ratio ventilation
Tidal VolumeSetting Range: 50~1500ML Display Range: 0~2000ml ; resolution ratio: 10ml
Inspiratory Plateau:0~1s
Maximum Safety Pressure≤ 6.0kPa
Sigh:one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
Pressure limit range:1.0~ 6.0kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicatingLower: 0.1 kPa ~ 6.0 kPa; Upper: 0 ~ 5.0 kPa
±0.2 kPa
Ventilation Volume alarm:≥ 18L/ MIN
Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
Operating conditions
Ambient temperature:10 ~ 40oC
Relative humidity:no higher than 80 %
Atmospheric pressure:860 hPa ~ 1060 hPa
Power requirement:100-120 Vac, 50/60 Hz;
Attention: the AC power supply used for the anesthesia machine must be well grounded.
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.
Storage
Ambient temperature:-15oC ~ +50oC
Relative humidity:no higher than 95 %
Atmospheric pressure:86 kPa ~ 106 kPa.
 It should be stored in a room without corrosive gas and well ventilated
Package
packaging boxcomply with the requirement of GB/T 15464
 Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation
 Moisture protection and rain protection to ensure that product is protected from natural damage.
Safty & Alarm
Gas source pressureImmediate alarm; “high level alarm” indicator flickers raising an audible alarm.
Airway Pressure AlarmThe airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The “high level alarm” indicator flickers raising an audible alarm.
Tidal Volume AlarmLower than 50ml, The alarm will be raised after one breathing cycle.
Ventilation Volume AlarmLower: 0~12L/Min; Up: 18L/ Min
Battery Voltage AlarmIt alrms when the battery voltage is lower than 10.5V.
Power On Self TestSelf test before working, assure it's working normally.

China First Aid Equipment Real Time Waveform Display Ventilator With 10.4 Inch LCD Display supplier

First Aid Equipment Real Time Waveform Display Ventilator With 10.4 Inch LCD Display

Inquiry Cart 0