Electrically Controlled Pneumatic Ventilator , Portable Ventilator For Ambulance

Brand Name:MeCan
Certification:ISO
Model Number:MC-500
Minimum Order Quantity:1 set
Delivery Time:3-15 working dates
Payment Terms:L/C, D/A, D/P, T/T, Western Union, MoneyGram
Contact Now

Add to Cart

Active Member
Location: Guangzhou Guangdong China
Address: RM718, EASTERN BUILDING, XIAOBEI, YUEXIU, GUANGZHOU, CHINA.
Supplier`s last login times: within 36 hours
Product Details Company Profile
Product Details

CE Approved Electrically Controlled Pneumatic Ventilator Portable ICU Ventilator


Model : MC-500


Prospective service and support


The MC-500 ventilator is an electrically controlled pneumatic ventilator integrating such functions as time, volume cycling, pressure limit, etc. It is mainly intended for providing ventilation support to a critically ill patient during the life threatening phase and ensuring the going-through of the dangerous period by the patient and smooth treatment of primary diseases for recovery. Also it provides an alternation in case of irreversible lesions in respiratory muscles or irreversible damage to upper airway to maintain the respiratory function of the patient, and also provides ventilation assistance for the patient during the recovery from a disease or operation.


Display: LED Display Following Parameters


Operating mode and respiratory rate;


——Ventilator control frequency set by medical staff;


——Tidal volume of each breath in the patient;


——Occurrence of spontaneous breath in the patient and spontaneous breathing frequency;


——Expiratory phase and inspiratory phase and actual respiratory rate;


——Minute volume;


——Setting and adjustment of inspiration triggering pressure, PEEP, and highest airway pressure;


Portable Ventilator Construction :


--Environmental Protection Material construction, hardly to be broken.


-- Trolly: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions


TECHNIQUES SPECIFICATIONS
Physical specifications
DisplayLED Display
Mode:pneumatically powered and electrically controlled system
Trolley:Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Technical Specifications:
Gas requirement:Medical oxygen with a pressure ranging from 280~600kpa
safety valve<12.5 kPa
Respiratory Rate6~60 times/ min
Oxygen concentration output<50%
minute volume≥18L
modes of ventilationA/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL
PEEP:0.1 ~ 1.0 kPa
Ptr:-4 ~ 10 hPa
IMV frequency:1~12 bpm
I/E Ratio:1:1.5; 1:2.0; 1:2.5; 1:3.0
Tidal Volume0-1500ml
Inspiratory Plateau:0~1s
Maximum Safety Pressure≤ 6.0kPa
Sigh:one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
Pressure limit range:0 ~ 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicatingLower: 0.9 kPa ~ 5.4 kPa; Upper: 0.5 kPa
±0.2 kPa
Ventilation Volume alarm:≥ 18L/ MIN
Power Supply AlarmAc/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
Operating conditions
Ambient temperature:10 ~ 40oC
Relative humidity:no higher than 80 %
Atmospheric pressure:860 hPa ~ 1060 hPa
Power requirement:100-120 Vac, 50/60 Hz;
Attention: the AC power supply used for the anesthesia machine must be well grounded.
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.
Storage
Ambient temperature:-15oC ~ +50oC
Relative humidity:no higher than 95 %
Atmospheric pressure:86 kPa ~ 106 kPa.
 It should be stored in a room without corrosive gas and well ventilated
Package
packaging boxcomply with the requirement of GB/T 15464
 Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation
 Moisture protection and rain protection to ensure that product is protected from natural damage.
Safty & Alarm
Gas source pressureImmediate alarm; “high level alarm” indicator flickers raising an audible alarm.
Airway Pressure AlarmThe airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The “high level alarm” indicator flickers raising an audible alarm.
Tidal Volume AlarmLower than 50ml, The alarm will be raised after one breathing cycle.
Ventilation Volume AlarmLower: 0~12L/Min; Up: 18L/ Min
Battery Voltage AlarmIt alrms when the battery voltage is lower than 10.5V.
Power On Self TestSelf test before working, assure it's working normally.

China Electrically Controlled Pneumatic Ventilator , Portable Ventilator For Ambulance supplier

Electrically Controlled Pneumatic Ventilator , Portable Ventilator For Ambulance

Inquiry Cart 0