Add to Cart
Cold Balanced Control System Programmable Temperature and Humidity Environmental Test Chamber
Technical Data Download:
iqualitrol Vertical Temperature Humidity Test Chamber
Brief Introduction:
Cold Balanced Control System Programmable Temperature and Humidity
Environmental Test Chamber was used to test the effects of
specified environmental conditions on biological items, industrial
products, materials, and electronic devices and components.
Technical Specifications:
Control Mode | Cold Balanced Control System |
Operating Temp. & Humid. Range | +5deg ~ +35deg; <85%R.H. |
Temp. range | 0℃ - 150℃ |
Humid. range | 10% ~98%R.H. |
Temp. & Humid. fluctuation | ±0.5deg; ±1.5%R.H. |
Temp. & Humid. uniformity | ≤2.0deg; ≤3%R.H. |
Temp. & Humid. Deviation | ≤1.0deg; ≤2%R.H. |
Temp. Heating Time | 0deg ~ +100degwithin 30min; -20deg ~ +150degwithin 45min |
Temp. Cooling Time | 20deg ~ -40degwithin 60min; 20deg ~ -70degwithin 85min |
Power Supply | AC380V, 50/60HZ, Three-phase (Specified by User) |
Exterior Material | stainless steel with baking paint |
Interior material | SUS304# Stainless steel |
Insulation material | Rigid Polyurethane foam |
Refrigeration system | Mechanical cascade refrigeration system; Fin type radiator |
Circulation system | Mechanical convection system |
Humidification Water Supply | Automatic water regulating, recoverable supply system, |
Water quality | Distilled water only, 20L Water tank capacity |
Controller | Touch Screen Controller |
Safety devices | Overheat protector Switch, Compressor overload protector Switch, water shortage protector Switch, Humidifier protector Switch, Fault alarm system |
Model Selection:
Model | Temp and Humidity Range | Internal Dimension (W x H x D) mm | Exterior Dimension (W x H x D) mm |
QTH-600A | 0℃~﹢150℃ | 700*1300*700mm | 950*1900*1300mm |
Remark: with dehumidifier, possible to work at 10℃ 20% to 80%RH, 5 shelfs + door lock | |||
Support custom made dimension |
Main Features:
1, Chamber exterior material is stainless steel with
environmentally baking paint resists corrosion and provides impact
resistance. Interior material is SUS304# stainless steel with
excellent heat resistance and easy to clean.
2, Touch screen controller is designed to save chamber programming
and setup time with temperature limit and alarm to protect your
product.
3, Safety relay connection is provided to protect your device under
test by removing power to it when the chamber is not running.
4, RS-232 communications is for computer connection, programming
can be set on computer by software, monitor testing process and
automatically execute power on/off functions.
5, Fog-free viewing window and interior light makes viewing
workspace freely and observe the test under best conditions.
6, Adjustable product shelf slides out for easier product access.
Shelf design is non-tipping and supports large product loads.
7, Left side of chamber with diameter 50mm (support customized)
cable port for power-on test.
8, Optional electronic humidity sensor is used on all test chambers
for accuracy and minimal maintenance.
Optional Accessories:
1, Cable Port: Size of cable port is available for 100mm, standard is 50mm;
2, Water purifier RO 80: Continuously provide purified water for humidifying heater and
wet-bulb;
3, Dehumidifier: The rotation regenerating dehumidifier M-120 ensures precise
control of low humidity (10C, 15%RH) for electrostatic reliability
tests;
4, Defrosting Device: The chamber automatically detects and melts the frost on the
evaporator when operating below 0deg in order to allow continuous
operation;
5, Electronic humidity sensor: Precision Humidity Sensor with stainless steel protection tube.