EJA110A-DMS5A-22ND Yokogawa EJA110A Differential Pressure
Transmitter
Ordering Information
Model: EJA110A - Differential pressure transmitter
Output Signal: -D - 4 to 20 mA DC with digital communication (BRAIN protocol)
Measurement span (capsule): M - 1 to 100 kPa {100 to 10000 mmH2O} {4 to 400 inH2O} {10 to 1000
mbar}
Wetted parts material?: S# - [Body]2 SCS14A; SUS316L3; [Vent plug] SUS3164
Process connections: *5 - without process connector (1/4 NPT female on the cover flanges)
Bolts and nuts material: *A - SCM435: [Maximum working pressure] 16 MPa {160 kgf/cm2}5
Installation: -2 - Vertical impulse piping type, right side high pressure, process
connector upside6
Electrical connection: *2 - 1/2 NPT female, two electrical connections without blind plug
Integral indicator: *N - (None)
Mounting bracket: D - SUS304: 2-inch pipe mounting (L type)
NOTE:
The '*' marks indicate the most typical selection for each
specification.
The '#' marks indicate the construction materials conform to NACE
material recommendations per MR01-75. For the use of SUS316
material, there may be certain limitations for pressure and
temperature. Please refer to NACE standards for details.
- Users must consider the characteristics of selected wetted parts
material and the influence of process fluids. The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also possible that the
diaphragm itself can be damaged and that material from the broken diaphragm and the fill fluid can contaminate the user's
process fluids. Be very careful with highly corrosive process fluids such as
hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature steam. Contact Yokogawa for detailed information of the wetted parts material.
- Indicates material of cover flanges and process connectors.
- Diaphragm material is Hastelloy C-276 or ASTM N10276. Other capsule
wetted parts materials are SUSF316L, SUS316L or ASTM grade 316L.
- SUS316 or ASTM grade 316.
- For Capsule code L when combined with Wetted parts material code H,
M, T, A, D, and B, the maximum working pressure is 3.5 MPa {35 kgf/cm2}.
- If necessary, specify Mounting bracket code C, D or K.
The EJA-A series is Yokogawa's most successful pressure transmitter
line. First released in 1991, it continues to offer high
performance and high reliability for any application. With an
installed base of over 4-1/2 million EJA-A transmitters worldwide,
it
has a proven track record that makes it a true industry workhorse.
The EJA110A is the Traditional-mount Differential Pressure
Transmitters in the series.
EJA110A features include:
- 0.065% Accuracy
- 0.1% Stability per 5 years
- 2,300 psi MWP
Model Number: EJA110A
Output: 4-20mA with HART protocol
Accuracy: ±0.065% of span
Cover Flange: SUS316, Hastelloy C-276, Monel
Power Supply: 10.5 to 42V
EJA110A Overview
Measurement Types |
Primary Variable | Differential Pressure (DP) |
Reference Accuracy |
Primary Variable | ±0.065% of DP Span |
Long Term Stability (All Normal Operating Conditions) |
Primary Variable | M, H, or V Capsule: ±0.1% of URL per 5 years L Capsule: ±0.1% of URL for 1 year |
Maximum Pressure Limit (MWP) |
All Capsules | 2,300 psi |
Over-pressure Effect |
Primary Variable | ±0.03% of URL |
Rangeability |
Primary Variable | M, H, or V Capsule: 100:1 L Capsule: 20:1 |
Burst Pressure (Absolute) |
All Capsules | 10,000 osi |