Add to Cart
Product Description:
Low Range NO 5NFL Nitric Oxide Sensor Product Code: AF508-W0P
Features:
The purpose of this document is to present the performance specification of the 5NFL nitric oxide gas sensor.
This document should be used in conjunction with Operating Principles (OP20) and the Product Safety Datasheet (PSDS 18).
The data provided in this document are valid at 20°C, 50% RH and 1013 mBar for 3 months from the date of sensor manufacture.
Output signal can drift below the lower limit over time. For guidance on the safe use of the sensor, please refer to the Operating Principles (OP20).
Cross Sensitivity Table
Whilst CiTiceLs are designed to be highly specific to the gas they are intended to measure, they will still respond to some degree to various other gases. The table below is not exclusive and other gases not included in the table may still cause a sensor to react.
IMPORTANT NOTE : The cross sensitivity data shown below does not form part of the product specification and is supplied for guidance only. Values quoted are based on tests conducted on a small number of sensors and any batch may show significant variation. For the most accurate measurements, an instrument should be calibrated using the gas under investigation.
Specifications:
Operating Principle | 4-electrode electrochemical |
Measurement Range | 0 ppm to 300 ppm NO |
Maximum Overload | 1000 ppm NO |
Filter | To remove effect of SO2 |