Display SpO2, PR | Battery Type: 3.7v (1000mAh) Rechargeable Lithium Batteries |
Pulse Rate: Measurement range: 25 - 250bpm Accuracy: ±2bpm | SpO2: Measurement range: 35% - 100% Accuracy: ±2% (80% - 100%) ;±3% (70% - 79%) |
Working Voltage: DC 3.4v ~ DC 4.3v | Free App: “BerrySmartHealth” |
Size:68(L)x58(W)x20(D)mm | Weight: 50g |
Comparison: Reference Pulse Oximeter | ||||||||||
Berry Pulse Oximeter | 70 ~100 | 90 ~ 100 | 80 - < 90 | 67 - < 80 | RESE≤3% Within the range of 70~100% | |||||
# pts | 240 | 79 | 81 | 81 | ||||||
Error | -0.6 | -1.1 | -0.9 | 0.1 | Pass | |||||
Root mean square error | 1.9 | 1.7 | 2.0 | 2.1 |
Mean Blood Oxygen | ||||
Average blood oxygen calculation accuracy: 98.51% | ||||
Diagnosis of hypoxemia | ||||
Sensitivity: 89.29% | Specificity: 77.78% | Accuracy:92.59% | ||
OSAHS Judgment Results | ||||
Sensitivity: 94.17% | Specificity: 61.21% | Accuracy:89.62% |
Display SpO2, PR | Battery Type: 3.7v (1000mAh) Rechargeable Lithium Batteries |
Pulse Rate: Measurement range: 25 - 250bpm Accuracy: ±2bpm | SpO2: Measurement range: 35% - 100% Accuracy: ±2% (80% - 100%) ;±3% (70% - 79%) |
Working Voltage: DC 3.4v ~ DC 4.3v | Free App: “BerrySmartHealth” |
Size:68(L)x58(W)x20(D)mm | Weight: 50g |
Comparison: Reference Pulse Oximeter | ||||||||||
Berry Pulse Oximeter | 70 ~100 | 90 ~ 100 | 80 - < 90 | 67 - < 80 | RESE≤3% Within the range of 70~100% | |||||
# pts | 240 | 79 | 81 | 81 | ||||||
Error | -0.6 | -1.1 | -0.9 | 0.1 | Pass | |||||
Root mean square error | 1.9 | 1.7 | 2.0 | 2.1 |
Mean Blood Oxygen | ||||
Average blood oxygen calculation accuracy: 98.51% | ||||
Diagnosis of hypoxemia | ||||
Sensitivity: 89.29% | Specificity: 77.78% | Accuracy:92.59% | ||
OSAHS Judgment Results | ||||
Sensitivity: 94.17% | Specificity: 61.21% | Accuracy:89.62% |