Add to Cart
Coordinate Measuring Machine Athena Series
Product Introduction:
From the base machine structure point of view, Athena features all-aluminum construction and utilized comprehensive CAD techniques, including Finite Element Analysis (FEA) and Module Analysis, with the result of improved thermal, dimensional/torsional, and geometrical stability and reduces structural deflection and the effects of vibrations and thermal variation on measuring performance. Wrap-around air bearing design optimizes the axis kinematics movement. Therefore, the whole system characterizes the optimum measuring repeatability and long-term system consistency.
Product Characteristic:
Specification:
Model | Maximum Permissible Error MPE(μm), L(mm), (ISO10360-2) | 3D Max. Speed (mm/s) | 3D Max. Acceleration (mm/s²) | ||||||
HP-TMe | HP-THDe | HP-S-X1 | |||||||
MPEE | MPEP | MPEE | MPEP | MPEE | MPEP | MPETHP/T | |||
Athena 564 | 2.3+L/300 | 2.5 | 2.1+L/300 | 2.3 | 1.9+L/300 | 2.1 | 3.0/68 | 520 | 1730 |
Athena 686 | 2.3+L/300 | 2.5 | 2.1+L/300 | 2.3 | 1.9+L/300 | 2.1 | 3.0/68 | 520 | 1730 |
Athena 8106 | 2.4+L/300 | 2.6 | 2.2+L/300 | 2.5 | 2.0+L/300 | 2.2 | 3.5/68 | 520 | 1730 |
Athena 8126 | 2.4+L/300 | 2.6 | 2.2+L/300 | 2.5 | 2.0+L/300 | 2.2 | 3.5/68 | 520 | 1730 |
Athena 8156 | 2.4+L/300 | 2.6 | 2.2+L/300 | 2.5 | 2.0+L/300 | 2.2 | 3.5/68 | 520 | 1730 |
Athena 10128 | 2.6+L/300 | 2.8 | 2.4+L/300 | 2.7 | 2.3+L/300 | 2.5 | 4.0/68 | 520 | 1730 |
Athena 10158 | 2.6+L/300 | 2.8 | 2.4+L/300 | 2.7 | 2.3+L/300 | 2.5 | 4.0/68 | 520 | 1730 |
Athena 10218 | 2.6+L/300 | 2.8 | 2.4+L/300 | 2.7 | 2.3+L/300 | 2.5 | 4.0/68 | 520 | 1730 |
Athena 10308 | 2.6+L/300 | 2.8 | 2.4+L/300 | 2.7 | 2.3+L/300 | 2.5 | 4.0/68 | 520 | 1730 |
Athena 121510 | 2.9+L/300 | 3.4 | 2.7+L/300 | 3.3 | 2.6+L/300 | 3.1 | 5.0/68 | 433 | 1040 |
Athena 122210 | 2.9+L/300 | 3.4 | 2.7+L/300 | 3.3 | 2.6+L/300 | 3.1 | 5.0/68 | 433 | 1040 |
Athena 123010 | 2.9+L/300 | 3.4 | 2.7+L/300 | 3.3 | 2.6+L/300 | 3.1 | 5.0/68 | 433 | 1040 |
Athena 152210 | 3.4+L/300 | 3.9 | 3.2+L/300 | 3.8 | 3.1+L/300 | 3.6 | 5.2/68 | 433 | 1040 |
Athena 153010 | 3.4+L/300 | 3.9 | 3.2+L/300 | 3.8 | 3.1+L/300 | 3.6 | 5.2/68 | 433 | 1040 |
Athena 152212 | 3.9+L/300 | 4.4 | 3.7+L/300 | 4.3 | 3.6+L/300 | 4.1 | 5.2/68 | 433 | 1040 |
Athena 153012 | 3.9+L/300 | 4.4 | 3.7+L/300 | 4.3 | 3.6+L/300 | 4.1 | 5.2/68 | 433 | 1040 |
Model | Range of Travel(mm) | Dimensions(mm) | Daylights(mm) | Maximum weight of the measured workpiece (Kg) | Machine Weight (Kg) | ||||||
X | Y | Z | Lx | Ly | Lz | Dx | Dz | Dz1 | |||
Athena 564 | 500 | 600 | 400 | 1055 | 1535 | 2247 | 634 | 114 | 594 | 300 | 590 |
Athena 686 | 600 | 800 | 600 | 1150 | 1735 | 2630 | 734 | 114 | 794 | 300 | 730 |
Athena 8106 | 800 | 1000 | 600 | 1350 | 1935 | 2640 | 934 | 114 | 794 | 500 | 1074 |
Athena 8126 | 800 | 1200 | 600 | 1350 | 2135 | 2640 | 934 | 114 | 794 | 500 | 1196 |
Athena 8156 | 800 | 1500 | 600 | 1350 | 2435 | 2640 | 934 | 114 | 794 | 500 | 1379 |
Athena 10128 | 1000 | 1200 | 800 | 1615 | 2220 | 2940 | 1130 | 118 | 940 | 1300 | 1785 |
Athena 10158 | 1000 | 1500 | 800 | 1615 | 2520 | 2940 | 1130 | 118 | 940 | 1500 | 2090 |
Athena 10218 | 1000 | 2100 | 800 | 1615 | 3120 | 2950 | 1130 | 118 | 940 | 1800 | 2625 |
Athena 10308 | 1000 | 2980 | 800 | 1615 | 4020 | 3010 | 1130 | 118 | 940 | 2000 | 4400 |
Athena 121510 | 1200 | 1500 | 1000 | 1840 | 2895 | 3360 | 1339 | 150 | 1173 | 1800 | 3792 |
Athena 122210 | 1200 | 2200 | 1000 | 1840 | 3595 | 3410 | 1339 | 150 | 1173 | 2250 | 5696 |
Athena 123010 | 1200 | 3000 | 1000 | 1840 | 4395 | 3440 | 1339 | 150 | 1173 | 2250 | 7637 |
Athena 152210 | 1500 | 2200 | 1000 | 2140 | 3595 | 3410 | 1639 | 150 | 1173 | 2250 | 6730 |
Athena 153010 | 1500 | 3000 | 1000 | 2140 | 4395 | 3440 | 1639 | 150 | 1173 | 2250 | 9046 |
Athena 152212 | 1500 | 2200 | 1200 | 2140 | 3595 | 3810 | 1639 | 150 | 1373 | 2250 | 6760 |
Athena 153012 | 1500 | 3000 | 1200 | 2140 | 4395 | 3840 | 1639 | 150 | 1373 | 2250 | 9076 |
Probe configuration for performance testing: HP-TMe/HP-THDe: Standard force measurement, The probe is 10mm long,The tip diameter is 4 mm. HP-S-X1:HP-S-X1S/HP-S-X1H/HP-S-X1C, The probe is 50mm long,The tip diameter is 5 mm. Performance metrics are valid under the following conditions: Standard ambient temperature: 18 - 22°C Ambient temperature changes : 1°C/h - 2°C/24h Temperature gradient: 1°C/m Relative humidity: 25% - 75% |
RationalDMIS2 Measuring Software:
It has obtained ANSI, ISO, GB, PTB standard certification, always synchronized with advanced standards and advanced design concepts, and fully supports the internationally popular I++ control protocol.It has a user-friendly and simple software interface, unique drag-and-drop quick operation, and seamless connection with CAD data. 100% graphical visualization from measurement to output report, which provides a complete measurement solution for enterprises.
Rational DMIS software integrates CAD, thin-walled, and scanning modules to meet the growing measurement needs of our customers.
Features of Rational DMIS:
● Supports IGES, STEP, DXF, STL, XYZ and other formats;
● Supports CATIA, UG, PRO-E, Solidworks, Parasolid direct reading interface;
● Intelligent Detection: real-time simulation of the detection path, automatic optimisation of the detection path;
● Complete and fast tolerance calculation: in line with GB, ISO, Y14.5, ANSI, DIN, AGMA and other standards;
● Diverse report output; Support HTML, Excel, PDF, OUT, TXT format;
● Abundant special modules: SPC statistical analysis, CCD composite image, line laser scanning, blade analysis, cylindrical involute gear analysis, cam analysis, network programming, network report;
● The core algorithm is certified by the Physikalisch-Technischen Bundesanstalt (PTB).