Add to Cart
Function | Range | Resolution | Accuracy |
AC volts[1] | 6.000 V | 0.001 V | 1.0 % + 3 (45 Hz to 500 Hz) |
60.00 V | 0.01 V | ||
600.0 V | 0.1 V | 2.0 % + 3 (500 Hz to 1 kHz) | |
1000 V | 1 V | ||
DC volts | 6.000 V | 0.001 V | 0.09 % + 3 |
60.00 V | 0.01 V | ||
600.0 V | 0.1 V | ||
1500 V | 1 V | 0.15 % + 2 | |
AC mV[1] | 600.0 mV | 0.1 mV | 1.0% + 3 (45 Hz ~ 500 Hz) 2.0% + 3 (500 Hz ~ 1000 Hz) |
DC mV | 600.0 mV | 0.1 mV | 0.09 % + 2 |
Continuity | 600 Ω | 1 Ω | |
Resistance | 600.0 Ω | 0.1 Ω | 0.5 % + 4 |
6.000 kΩ | 0.001 kΩ | ||
60.00 kΩ | 0.01 kΩ | ||
600.0 kΩ | 0.1 kΩ | ||
6.000 MΩ | 0.001 MΩ | ||
50.00 MΩ | 0.01 MΩ | 1.5 % + 4 | |
Capacitance | 1000 nF | 1 nF | 1.2 % + 2 |
10.00 μF | 0.01 μF | ||
100.0 μF | 0.1 μF | ||
9999 μF | 1 μF | 10 % typical | |
Frequency | 99.99 Hz | 0.01 Hz | 0.1 % + 2 |
999.9 Hz | 0.1 Hz | ||
9.999 kHz | 0.001 kHz | ||
99.99 kHz | 0.01 kHz | ||
AC VA (45~500Hz)[1] | 360.0 VA | 0.1 VA | 2 % + 1.0 VA |
3.600 kVA | 0.001 kVA | 2 % + 0.01 kVA | |
36.00 kVA | 0.01 kVA | 2 % + 0.1 kVA | |
60.00 kVA | 0.01 kVA | 2 % + 0.15 kVA | |
DC VA[1] | 360.0 VA | 0.1 VA | 2 % + 1.0 VA |
3.600 kVA | 0.001 kVA | 2 % + 0.01 kVA | |
36.00 kVA | 0.01 kVA | 2 % + 0.1 kVA | |
90.00 kVA | 0.01 kVA | 2 % + 0.25 kVA | |
Min-max avg | For DC functions, accuracy is the specified accuracy of the measurement function ± 12 counts for changes longer than 350 ms in duration. | ||
For AC functions, accuracy is the specified accuracy of the measurement function ± 40 counts for changes longer than 900 ms in duration. |
[1] <1% range, unspecified
Environmental specifications | |
Operating temperature | -10 °C to 60 °C |
Storage temperature | -30 °C to +70 °C |
Humidity (without condensation) | 0 % to 90 %, 10 °C to 30 °C |
0 % to 75 %, 30 °C to 40 °C | |
0 % to 45 %, 40 °C to 60 °C |
Warranty and Protection | |
Electromagnetic compatibility | IEC 61326-1, IEC 61326-2-2, Portable, Group 1, Class A |
Overvoltage category | CAT III 1500 V, CAT IV 1000 V |
Agency approvals | ETL (AMER and EMEA), CSA (APAC), CE, UK CA |
Drop test | 2 meter (6.5 ft) drop test |
Ingress protection | IP52 |
Warranty | Limited lifetime |
Mechanical and general specifications | |
Counts | 6000 |
Size | 22.5 cm x 10.5 cm x 5.7 cm (8.8 in x 4.1 in x 2.2 in) |
Weight | 0.7 kg (1.5 lb) |
Battery life | > 150 hours typical, without backlight (Alkaline, 3 AA) |
> 100 hours typical when connected to wireless current clamp (Alkaline, 3 AA) |
Function | Range | Resolution | Accuracy |
AC current (True-RMS)[2] | 60 A | 0.01 A | 1.5% + 0.15 A, 45~500Hz |
DC current[2] | 60 A | 0.01 A | 1.5% + 0.15 A |
General specifications – Fluke a283 FC Wireless Current Clamp | |
Operating temperature | -10 °C to 60 °C |
Storage temperature | -30 °C to +70 °C |
Humidity (without condensation) | 0 % to 90 %, 5 °C to 30 °C |
0 % to 75 %, 30 °C to 40 °C | |
0 % to 45 %, 40 °C to 60 °C |
Warranty and Protection | |
Electromagnetic compatibility | IEC 61326-1, IEC 61326-2-2, Portable, Group 1, Class A |
Overvoltage category | CAT III 1500 V, CAT IV 1000 V |
Agency approvals | ETL (AMER and EMEA), CSA (APAC), CE, UK CA |
Drop test | 2 meter (6.5 ft) drop test |
Ingress protection | IP52 |
Warranty | 1-year |
Mechanical and general specifications | |
Size | 226.0 mm x 91.0 mm x 42.0 mm (8.9 in x 3.6 in x 1.7 in) |
Jaw opening | 34 mm (1.34 in) |
Weight | 0.375 kg (13.2 oz) |
Battery life | > 80 hours typical (Alkaline, 2 AAA) |
Features | |
LED indicator | Indicates wireless connection and data transmission status |
Battery indicator | Indicates battery status |
Power Key | Power On/Off |
Zero Key | A DC zeroing function |
[2] <1% range, unspecified