Add to Cart
Multipurpose 02 Sensor Medical,Practical MOX4 Oxygen Sensor
MOX4 OXYGEN SENSOR
Operating Principle | Partial pressure electrochemical |
Output | 9 mV to 13 mV in 210 mBar O2 |
Measurement Range | 0 mBar to 1500 mBar O2 |
Response Time | <15 s (air to 100% O2 ) |
Baseline Offset | <200 μV |
Linearity | Linear 0% to 100% O2 |
Temperature Compensation | Partial compensation for Teledyne T-7 equivalence |
External Load Resistor | 10 kΩ min |
Connector | UK Data Socket AMP (P/N 520257-2) |
Housing Material | White ABS |
Typical Applications | Critical care anesthesia |
Operating Temperature Range | -20°C to +50°C |
Operating Pressure Range | 0.5 Bar to 2.0 Bar |
Operating Humidity Range | 0% to 99% RH noncondensing |
Long-Term Ouput Drift in 100% O2 | < 5% signal loss/year |
Recommended Storage Temperature | -10°C to 40°C (short excursions to 50°C allowed) |
Expected Operating Life: @ 20°C @ 40°C | 1.5 x 106 % O2 hours 0.8 x 106 % O2 hours |
Delivery Time:
Shipment and Payment:
Customization and OEM Service:
Distributor Policy: