Add to Cart
Biosafety Cabinet Class II / Biosafety Cabinet Company / Biosafety
Cabinet Clean Room Equipment
Features of Biosafety Cabinet Class A2:
1.Front window: when using optical perspective is clear, clean and
disinfect does not negatively
impact burst-proof tempered glass, tempered glass production, glass
thickness ≥ 5 mm.
2.Cabinet: four two-tier structure, all Studio and the collecting
tank to use imports 304 stainless
steel materials.
3.Operations on three sides and side wall panels unibody structure,
leave some space
at the bottom,easy to clean, to avoid cross infection.
4.Blower/motor maintenance and disassembly, replacement HEPA
filters are available from all
safety cabinets.
5.Containment of the Cabinet: the cabinets pressurized to 500Pa,
keep the air pressure
is not lower than 450Pa after 30min.
6.Cabinet stability: a countertop Center load 23kg unload,
countertop must not bend downward.
7.Alarm system: safety Cabinet front window opening height exceeds
a set level, decreased
safety Cabinet flow velocity and flow fluctuation of air flow velocity in
excess of their nominal
value of 20%, the interlocking system startup, sound, light, sound and
alarm.
8.HEPA filter integrity: scan filter at any point of the leak rate
of less than 0.01%.
9.Cannot scan filters detect leak rate of less than 0.005% personal
protection: percussive
sampler number ≤ 10CFU/.
10.Slit type sampler number ≤ 5CFU/product protection: colony count
≤ 5CFU/cross
infection: colony count ≤ 2CFU/.
Technical Parameters of Biosafety Cabinet Class A2:
Product name | Biosafety Cabinet Class A2 | ||
Index/type technical parameter | HZJBSC-1100-LⅡA2 | HZJBSC-1360-LⅡA2 | HZJBSC-1800IIA2 |
Exhaust volume | 30% exhaust | ||
HEPA filter leakage ratio | ≤0.005% | ||
Wind speed | Inflows flow average wind speed: >0.5M/s | ||
Noise | ≤62dB(A) | ||
Semi-peak | ≤3μm(rms) | ||
Illumination | ≥700LX | ||
Display | LCD + remote control | ||
Power supply | Single AC 220V/50Hz | ||
Normal power consumption | 0.85kw | ||
Max power consumption | 1.85kw (including two spare socket 1000W) | ||
Weight | 305kg | 320kg | 326kg |
Workspace dimensions | 1220x600x680mm | 1360x600x680mm | 1800x600x680mm |
Device overall dimension | 1382x790x2150mm | 1525x790x2150mm | 1940x790x2150mm |
Air filter specifications | 1215x500x50mmx① | 1355x500x50mmx① | 1795x500x50mmx① |
Exhaust air filter specifications | 600x440x90mmx① | 740x400x90mmx① | 865x440x90mmx① |
Fluorescent lights/UV lamps | 30w×②/30w×① | 40w×②/40w×① | 40w×②/40w×① |
The number of applicable | Single double single | Double single | More than |
Goods Informations: