The insulation fault locator AIL150-4/-8 in conjunction with AIM-M200 medical intelligent insulation monitor and ASG150 test signal generator, are designed for insulation fault location in medical IT systems.The locating current generated by ASG150 are detected using the integrated measuring current transformers and evaluated by the insulation fault locators.
Main Characteristics
● Insulation fault location in AC IT systems
● AIL150-4: 4 measuring channels
● AIL150-8: 8 measuring channels
● LED indicators: On, Comm
● CAN bus
● Auxiliary power supply: DC24V
Technical Parameters
Model
AIL150-4/8
Power Supply
Voltage
DC24V
Power consumption
≤2VA
Monitored System
Rated volatge
0~24V DC
Rated frequency
45~60Hz
Location Signal
Volatage
<25V
Current
<1mA
Fault Location
Maximum chanel
4/8
Response time
<2s
Communication
Interface
CAN bus
Protocol
Custom protocol

Dimension(mm)


Wiring