Acrel AIM-M300 Medical Intelligent Insulation Monitor Test Signal Generator Monitor De Aislamiento Para Hospitales

Brand Name:Acrel
Certification:CE
Model Number:AIM-M300
Minimum Order Quantity:1pc
Delivery Time:5-8 work days
Payment Terms:T/T
Contact Now

Add to Cart

Verified Supplier
Address: No. 9 Xincheng Road,Chengjiang Street, Jiangyin City, Jiangsu Province,China
Supplier`s last login times: within 33 hours
Product Details Company Profile
Product Details

AIM-M300 series medical intelligent insulation monitor


AIM-M300 series medical intelligent monitoring device adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is suitable for insulation monitoring of isolation power system in Class 2 medical locations 2 such as operating room and intensive care unit.


Function features of AIM-M300 series medical intelligent monitoring device:
 Functions of real-time monitoring and fault alarming of the grounding insulation resistance, transformer load current and transformer winding temperature of the monitored IT system;

 Locating current injector for insulation fault location systems;

 Real-time monitor the line disconnection fault, current sensor disconnectiobn,temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication within 2s after the fault occurs. when the fault occurs;
 Relay alarm output, LED alarm indication and other faults indication functions;

 Two kinds of fieldbus communication technology, which are used for centralized alarm and display device, insulation fault locator and upper computer management software communications, and monitoring the operation status of IT system in real time.

 Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel to analyze the operation conditions of system and promptly eliminate the faults.


Technical Parameters

AUX PowerVoltageDC 18-36VLocate SignalLocate voltage<12V
Power consumption≤6WLocate current≤1mA
Insulation MonitoringResistance range15-999kΩAlarm OutputOutput mode2 output relay
Response value50-999kΩContact ratingAC 250V/3A
DC 30V/3A
Relative uncertainty±10%Communication
Protocol
Communication 1RS485, Modbus-RTU
Response time≤3sCommunication 2CAN, customize
Permissible system
leakage capacitance
≤5μFEnvironmentOperating temperature-10~+55°C
Allowed external
voltage Ufg
≤DC 280VTransport temperature-25~+70°C
Measuring voltage≤13VStorage temperature-25~+70°C
Measuring current≤50μARelative humidity5-95%, No condensation
lmpedance Zi≥240kΩAltitude<2500m
Current MonitoringMeasuring Value2.1-60AInsulation methodguide
Measuring Accuracy±5%Display MethodLCD, LED
Alarm Value5-80AFault Record20
Temperature MonitoringTherrnal resistortwo Pt100IP degreeIP40
Measuring range50~+200°CRated impulse voltage/ pollution degree4KV/III
Alarm value range0~+200°CEMC/EMRIEC61326-2-4