Add to Cart
Three Electrode Eystem Lithium Ion Battery Analyzer Milliampere Tester
The Neware battery tester models listed below here in different categories are the most popular ones chosen by our customers , so if you belong to one of the following category , then it might be easier for you to find out the battery tester that meets your requirement based on your peers’ choice.
BTS-4000 series entered into market in the year of 2008 . Now it is one of the most successful and most popular battery testing system in China and World . Thousands of battery manufacturers , battery research institutes and laboratories use it as their primary testing partner for their various testing activities .
Battery researchers and scientists , battery manufactures , importers and exporters , you can always choose the most appropriate Neware battery testers for battery quality check , battery performance tests, battery material research , and battery mass production lines .
a . Neware 5V10mA / 20mA / 50mA is born for coin cell test , formerly its 5V1mA this modle . Students and researchers in universities and institutes just love this product , its the real star on their mind .
b . Testing with the uA current unit is more accurate and more convenient .
c . The lower range is able to output current low as 5uA , which can tell the subtle difference of various materials such electrodes , electrolyte , separator , etc .
d . Sensitive galvanometer is used for students to check weak current or tiny voltage in DC circuit in the lab . It is a kind of high sensitive magnetoelectric instrument , which can measure the tiny current .
e . It mainly focuses on battery research and development instituions and universities .
f . 20 set of mA Neware battery testers can be put in one rack
resolution | AD 16bit , DA 16bit | |
Input Power | 25W | |
Voltage | Per Channel Voltage Range | 25mV - 5V |
Discharge Min Voltage | 0V | |
Accuracy | 0.05 percent of FS | |
Stability | 0.05 percent of FS | |
Current | Per Channel Current Range | Range1 , 5uA – 1mA Range2 , 1mA – 25mA Range3 , 25mA – 50mA |
Accuracy | 0.05 percent of FS | |
Stability | 0.05 percent of FS | |
Power | Per Channel Output Power | 0.25W |
Stop current | Range 1 , 2uA Range 2 , 0.05mA Range 3 , 0.1mA | |
Stability | 0.1percent of FS | |
Data Recording | Record Condition | Min Current - Range 1 , 2uA Range 2 , 0.05mA Range 3 , 0.1mA |
Record Frequency | 10Hz | |
Charge | Charge Mode | CCCharge , CVCharge , CCCVCharge , CPCharge |
End Conditon | Voltage , Current , Relative Time , Capacity , -deltaV | |
Discharge | Discharge Mode | CCD , CPD , CRD |
End Condition | Voltage , Current , Relative Time , Capacity | |
Pulse | charge | CCC , CPC |
Discharge | CCD , CPD | |
Min Pulse Width | 500ms | |
Automated Switch | Automated swith from charge to discharge for each pulse | |
End Condition | Voltage, Test Time | |
Cycle | Loop Measure Range | 65535 times |
Max Steps Per Loop | 254 | |
Nested Loop | Nested loop Function , Max Support 3 Layers | |
Protection | Software Protection | Power - down Data Protection |
Off - line Testing Function | ||
Setting Protection Condition,Setting Parameter - Low Voltage limit , Upper Voltage limit , Low Current limit , Upper Current limit , Delay Time | ||
IP protection level | Protection level IP20 | |
Voltage and Current Testing Sample | 4 - wire connecting | |
Database | MySQL Database | |
A Means of Communication of Upper Machine | TCP / IP Protocol | |
Communication Interface | Ethernet Port |