6m3/h Gas Meter Test Bench Adopting Negative Pressure Method

Brand Name:HNMC
Model Number:MCGB001
Minimum Order Quantity:1
Delivery Time:4-8weeks
Payment Terms:Western Union, MoneyGram, L/C, T/T
Place of Origin:China
Contact Now

Add to Cart

Active Member
Location: Zhengzhou Henan China
Address: No.1207-1, Building #1, the National University Sci-Tech Park of Henan Province, No.11 Changchun Road, Zhengzhou, Henan 450000
Supplier`s last login times: within 38 hours
Product Details Company Profile
Product Details

0.016 m3/h~6 m3/h Gas Meter Test Bench (commercial gas meter)

Technical Proposal

Feature

  • Adopt Serial comparison method to test several different kinds of diaphragm gas meter.
  • Use VSBA Sonic nozzle to meet Qmax,0.2Qmax,Qmingas meter flow point measurement. It is convenient to change and clean nozzle regularly. (patent No.:20110237008.6)
  • Could accurately test temperature, pressure etc. and calculate error in time, guarantee device stable and accurate Ø
  • Adopt negative pressure method to ensure gas flow passed through to reach Critical flow state.
  • Test system could test automatically under control of PC software, meanwhile could save, print, analyze and statistic etc.
  • Also test manually gas meter without cursor .
  • graceful appearance, fine art-work,durable structure

0.016 m3/h~6 m3/h Gas Meter Test Bench (commercial gas meter)

Gas Test bench configuration unit and work principle

Gas Test bench is consist of compression gas source,vacuum bump, control gas circuit, test pipe system,VSBA Sonic nozzle, device running control system, error running system, operation platform and PC etc


Gas flow by vacuum pumping in test pipeline. When gas trough sonic nozzle reach critical flow condition,quality and flow of gas though tested meter and nozzle keep the same. Control and operation part work by electric. Mainly consist of error sampling calculation CPU,device running CPU and PC .

0.016 m3/h~6 m3/h Gas Meter Test Bench (commercial gas meter)

Main technical index and performance of 12 positions Sonic nozzle

Main technical index and performance of Sonic nozzle gas flow verification device (commercial gas meter)

Device flow range0.06 m3/h~40m3/h
Total uncertainty≤0.5% , k=2
Negative pressure
Test range of absolute pressure sensor0~120kPa,Accuracy ≤0.075%
Work condition
ambient temperature(20±5)℃
relative humidity(45~75)%RH
test range of temperature sensor(0—50)℃
accuracy≤±0.2℃
test range of humidity sensor0~100%RH
Gas meter SpecG6~G25
Installation methodclip meter by pneumatic method,test 1-2 nos meter(diaphragm meter) for serial test. Up to G10 for 2 positions,G25 or bigger is single position .Clip meter by volumetric structure only support flow gauge of DN25,DN40 meter. If it other diameter flow gauge, it will match other nozzle
Sampling methodcollect automatically or manual reading input, data dealing automatically, record save, certificate and test report print, data inquiry
Micro differential pressure sensor test range0~1000Pa,accuracy≤0.5%
XO frequency accuracy of time record<10-4,stability per day<10-5.Flow point:0.06m3/h,0.1 m3/h,0.25 m3/h ,0.4 m3/h ,1 m3/h,2 m3/h,4 m3/h,8 m3/h,16m3/h,32 m3/h,flow point deviation≤5%
Power supplyAC 3*380V±20V ,50Hz

Test pipeline system as below picture:

China 6m3/h Gas Meter Test Bench Adopting Negative Pressure Method supplier

6m3/h Gas Meter Test Bench Adopting Negative Pressure Method