Add to Cart
Bench Type LCR Digital Meters Inductance Measuring Instrument
VICTOR 4090C Bench Type Digital LCR meter Capacitance Resistance
Impedance Inductance Measure Instrument Electric Bridge Component
1.1 General Technical Specifications
●Power supply voltage: 220V AC ± 10%, or 110V AC ± 10%, 45-65Hz
●Working environment: 0 ~ 40 ℃, relative humidity <80%.
●Storage environment: -10 ~ 50 ℃, relative humidity <80%.
●Dimensions: 265mm * 105mm * 305mm (width * height* depth).
●Mass: 2.3kg.
●Power consumption: <10W
SN | Description |
1 | Power jack: AC 220V / 50Hz power supply input socket. |
2 | Voltage selector: 110V / 60Hz VAC or 220V / 50Hz VAC. |
3 | RS232 INTERFACE. |
4 | USB Device interface. |
The following data shall be measured under the following conditions:
C: 0.20%*(1+ Cx/Cmax+ Cmin/Cx )(1+Dx)(1+ks+kv+kf);
L: 0.20%*(1+ Lx/Lmax+ Lmin/Lx )(1+1/Qx)(1+ks+kv+kf);
Z: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+ks+kv+kf);
R: 0.20%*(1+ Rx/Rmax+ Rmin/Rx )(1+Qx)(1+ks+kv+kf);
D: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(1+Dx+Dx*Dx)(1+ks+kv+kf );
Q: 0.20%*(1+ Zx/Zmax+ Zmin/Zx )(Qx+1/Qx )(1+ks+kv+kf );
The maximum and minimum values of the measurement parameters that
affect accuracy
Frequency(Hz) | 100 | 120 | 200 | 400 | 800 | 1K | 2K | 4K | 8K | 10K |
Cmax | 800 | 667 | 400 | 200 | 100 | 80 | 40 | 20 | 10 | 8 |
Cmin | 1500 | 1250 | 750 | 375 | 187 | 150 | 75 | 37.5 | 18.7 | 15 |
Lmax | 1590 | 1325 | 795 | 397.5 | 198.8 | 159 | 79.5 | 39.8 | 19.9 | 15.9 |
Lmin | 3.2 | 2.6 | 1.6 | 0.8 | 0.4 | 0.32 | 0.16 | 0..08 | 0.04 | 0.032 |
Z/Rmax | 1 | |||||||||
Z/Rmin | 1.59 |
Frequency (Hz) | 15K | 20K | 40K | 50K | 80K | 100K |
Cmax | 5.3 | 4 | 2 | 1.6 | 1 | 0.8 |
Cmin | 10 | 7.5 | 3.75 | 3 | 1.87 | 1.5 |
Lmax | 10.6 | 7.95 | 3.98 | 3.18 | 1.99 | 1.59 |
Lmin | 0.021 | 0.016 | 0.008 | 0.0064 | 0.004 | 0.0032 |
Z/Rmax | 1 | |||||
Z/Rmin | 1.59 |
Wherein: the unit of Cmax is uF; the unit of Cmin is pF; the unit
of Lmax is H; the unit of Lmin is mH; the unit of Zmax / Rmax is
MΩ; the unit of Zmin / Rmin is in Ω;
Technical Data
Model | 4090A Type | 4090B Type | 4090C Type |
Digits | Main parameter: 5 digits; secondary parameter: 5 digits | ||
Measurement Parameter | Main parameter: L/C/R/Z; secondary parameter: X/D/Q/θ/ESR | ||
Measurement Range(L) | 0.01μH - 9999H | ||
Measurement Range(C) | 0.01pF - 99999μF | ||
Measurement Range(R) | 0.0001Ω - 99.99MΩ | ||
Basic Accuracy | 0.2% | ||
Test Frequency (Hz) | 10 point(100,120,200,400, | 12point(100,120,200, | 16point(100,120,200, |
Measurement Display Speed | 2 times/second (slow), 4 times/second (medium speed), 8 times/sec (fast) | ||
Internal Bias | 0-1500mV adjustable, 1mV stepping | ||
Test Level | 100mV,300mV,600mV,1000mV,1500mV,2000mV | ||
Correction Function | Open circuit correction, short circuit correction | ||
Screening function | The screening limit can be set from -50% to + 50% with fixed points of 1%, 5%, 10% and 20% | ||
Selection of comparator | Five gear sorting, third gear qualified, one gear failed, one ancillary gear | ||
Communication Interface | Standard configuration: USB, RS232 (or 485), Handle interface; Optional: GPIB, USB Host | ||
Others | Support DCR, electrolytic capacitor measurement mode, adjustable backlight brightness, optional in English and Chinese |