VICTOR 06 V/mA calibrator measure DCV DCA simulating output DC voltage, DC current
Input measurement function
[Applicable within one year after calibration, 23°C±5°C, 20~70%RH, Accuracy=± (%Setting Value+Digit)]
Measurement function | Range | Measuring range | Resolution | Accuracy | Remarks |
DC Voltage DCV | 100mV | -110.000mV~110.000mV | 0.001mV | 0.05%+0.03mV | Input resistance: Approximately 1MΩ |
1V | -1.1000V~1.1000V | 0.0001V | 0.05%+0.3mV | ||
35V | -35.000V~35.000V | 0.001V | 0.02%+2mV | ||
DC current DCI | 30mA | -35.000mA~35.000mA | 0.001mA | 0.02%+4uA | Shunt resistance: Approximately 10Ω Input resistance: Approximately 20Ω |
Loop power supply | 24 V | 10% |
Analog output function
[Applicable within one year after calibration, 23°C±5°C, 20~70%RH, Accuracy=±(%Setting Value+Digit)]
Output function | Range | Output range | Resolution | Accuracy | Remarks |
DC Voltage DCV | 10V | -1.000V~11.000V | 0.001V | 0.05%+2mV | Maximum output current: 5mA |
DC current DCI | 20mA | 0.000mA~22.000mA | 0.001mA | 0.05%+4uA | Max load of 1000Ω at 20 mA When used as an analog transmitter, the external loop power supply is 5~28V. |
Loop power supply | 24V | ±10% | Maximum output current 25mA |