VICTOR15+ Multifunction Process Calibrator Handheld Battery-Operated Instrument for Measuring and Sourcing Parameters with Pressure Calibration

Brand Name:VICTOR
Certification:CE
Model Number:VICTOR 15+
Minimum Order Quantity:1pcs
Delivery Time:5 to 30 days
Payment Terms:T/T, Western Union, L/C, Paypal
Contact Now

Add to Cart

Site Member
Address: JINGYUAN 7 ROAD, JINGHE INDUSTRIAL PARK, NORTH DISTRICT,XI'AN,CHINA  710200
Supplier`s last login times: within 22 hours
Product Details Company Profile
Product Details

VICTOR 15+ V/mA Calibrator current calibrator handheld calibrator
V/mA/Pressure Calibrator

  • Introduction

The voltage/current/pressure calibrator (the calibrator in the following) is a handheld, battery-operated instrument that measures and sources electrical and physical parameters.
Features:
Measure: DC-voltage, DC-current, frequency and continuity
Source: DC-voltage, DC-current, simulation transmitter, frequency and pulse

Pressure: measuring pressure, Calibrating pressure-voltage transmitter, Calibrating 2wire pressure transmitter, Calibrating pressure switch;
Others features:

  • Manual step source and auto -step and sweeping –step source
  • Measurement/source mA% display
  • Measurement wave-filter function
  • Measurement manual-holding function
  • Pressure source auto-holding function

Specifications
General Specifications for measure
These specifications assume:

  • A 1-year calibration cycle
  • An operating temperature of 18℃ to 28℃
  • Relative humidity of 35% to 70% (non_condensing)

Accuracy is expressed as ± (percentage of reading + percentage of range).

F FunctionReferenceRangeResolutionAccuracyRemark
DCV200mV

-20.00mV~
220.00mV

10μV0.02+0.02Input Resistance: 100MΩ
5V

-0.5000V~
5.5000V

0.1mV0.02+0.01Input Resistance: 1MΩ
50V

-5.000V~
55.000V

1mV0.03+0.01
DCmA50mA

-4.000mA~
55.000mA

1μA0.02+0.01Input Resistance: 5Ω
FREQ500Hz3Hz~500.00Hz0.01Hz±2digit

Input Impedance:100 kΩ at least;
Sensitivity:3Vp-p minimum;
Duty Cycle: 50%.

5KHz

3Hz~
5.0000KHz

0.1Hz
50KHz3Hz~50.000KHz1Hz
*CONT.≤250Ω soundOPEN/CLOSEOpen voltage: 2.5V


General Specifications for Source

FunctionReferenceRangeResolutionAccuracyRemark

DCV

1000mV-100.000mV~1100.000mV10μV0.02+0.01Maximum output current: 2mA
10V-1.0000V~11.0000V0.1mV0.02+0.01Maximum output current: 5mA

DCA

20mA0.000mA~22.000mA1μA0.02+0.02

External supply for simulate mA: 5V–28V
Maximum load 1KΩ at 20mA

FREQ100Hz1.00Hz~110.00Hz0.01Hz±2 count

Output voltage: 1~11 Vp-p (zero base waveform);
Amplitude accuracy: ±(10% +0.5V);
Maximum load: >100 KΩ;
Duty Cycle: 50%.

1KHz0.100kHz~1.100KHz1Hz
10KHz1.0KHz~11.0KHz0.1KHz
PULSE100Hz1~100000cycles1cyc_______

Output voltage: 1~11 Vp-p (zero base waveform);
Amplitude accuracy: ±(10% +0.5V);
Maximum load: >100 KΩ;
Duty Cycle: 50%.

1KHz
10KHz
LOOP24V±10%

Maximum current: 25 mA
Short circuit protected

Other feature:

l Temperature Coefficient: 0.1 times the applicable accuracy specification per degree ℃for 5℃ to 18℃ and 28℃ to 50℃.

l Maximum voltage between any output terminal and earth: 30Vpk

Maximum output current: Approximately 25mA.
General Specifications for Pressure

FunctionReferenceRangeResolutionAccuracyRemark
PRESSUREDetermined by pressure module_____Determined by pressure moduleDetermined by pressure moduleFor more detail, refer the pressure module about APM.

China VICTOR15+ Multifunction Process Calibrator Handheld Battery-Operated Instrument for Measuring and Sourcing Parameters with Pressure Calibration supplier