RS485 Water Quality Transmitter

Brand Name:Daruifuno
Certification:CE
Model Number:APF2-H3A
Minimum Order Quantity:Negotiable
Place of Origin:China
Price:Negotiable
Contact Now

Add to Cart

Verified Supplier
Location: Suzhou Jiangsu China
Address: Room 301, Building 3, No. 25 Datong Road, High-tech Zone, Suzhou City, JIANGSU P.R CHINA
Supplier`s last login times: within 1 hours
Product Details Company Profile
Product Details

RS485 Fluoride Ion Analyzer Ion Selective Electrode For Sewage Treatment Plants


Online fluoride ion analyzer is widely used in continuous monitoring of ion concentration in water discharged from sewage treatment plants, surface water, water plants and enterprises under various working conditions, in order to control the water quality to meet the prescribed water quality standards. The instrument uses ion selective electrode method to measure the corresponding potential difference corresponding to the fluoride ion concentration in the water. There is no need to add any reagents during the detection process and no secondary pollution.This meets the needs of environmental protection.


Advantages Of The Fluoride Analyzer


Support a variety of installation methods


Enclosure protection level IP66


Clock and historical data recording functions


A two-wire RS485 interface


128*64 large screen graphic dot matrix LCD screen


Set menu password to facilitate hierarchical management


Meet monitoring, dosing control and remote transmission of data signals at the same time



Parameters Of The Fluoride Analyzer


Product NameFluoride analyzer
Product Model

APF2-H3A

Software Version

DRFN ion electrode Analysis software V1.0

Measure Range(F-)

0.02~20mg/L

Measure Range(ORP)

±2000mV

Measure Range(TEMP)

-10~150℃

Resolution(F-)

0.01 mg/L

Resolution(ORP)

1mV

Resolution(TEMP)

0.1℃/ 0.1℉

Accuracy

3% of the measuring range

Stability

1% of the measuring range every 24 hours, no accumulation

Repeatability

Better than 1% of the measuring range

Relay Control

2 settable SPST relays, the max. load 3A/250VAC

Analog Current Output

2 settable 0/4~20mA active current loop output max. load 1000Ω

Communication Method

RS485, MODBUS RTU; support JSON text data format

Display Screen

128*64 LCD screen

Time History/Data Record

Year/month/day/hour/min/second, record 14000 historical data

Protection Grade

IP66

Operating Environment

0~60℃, relative humidity less than 95%

Storage Environment

-20~70℃, relative humidity less than 55%

Shell Material

Enhanced ABS

Dimensions

144*144*120mm (hole size 138*138 mm)

Mounting Method

Wall mounting, pipe clamping, panel installation

Power Supply

100~240V AC

Weight

800g

Power Consumption7W

China RS485 Water Quality Transmitter supplier

RS485 Water Quality Transmitter

Inquiry Cart 0