PHiliphh FM20 Fetal Monitor Paper Sensor Board Assembly M1353 66522 M2703-60003



►Brief Introduction

The paper sensor hardware consists of a reflective light sensor that detects the black marks on the trace paper, and paper-out. It is attached to the RFI Bracket, and connected to the Recorder Adapter Board via a removable cable connector. And paper recorder provides documentation of fetal traces and maternal parameters, together with user-defi ned annotations on an effi cient and easy to read documentation quality 15.2cm (6˝) paper strip. The recorder can also be set to run at high speed to record a 6 second maternal ECG strip in real time.


Specification

Option NumberDescription
Paper Scaling P0150 – 210 bpm paper scaling
Paper Scaling P0230 – 240 bpm paper scaling
Paper AdvanceUp to 20 mm/s

Recovering Traces on Paper

The monitor is able to recover traces by printing them out at a high speed from the monitor's backup memory. If the monitor runs out of paper, or if the paper drawer is open, the exact time when this happens is annotated in the backup memory. If the Bridge Paperout setting is set to On (default), when new paper is loaded and the recorder is started, a trace recovery printout of the data recovered from the backup memory is automatically printed out at high speed (up to 20 mm/s), starting from the time noted in the backup memory. This ensures that no data is lost. A minimum of one hour of trace recovery data can be printed out from the backup memory. When the trace recovery printout has finished, the recorder automatically switches back to continue recording the current trace at the normal speed.


► Removing the Paper Sensor Assembly

1. Remove the top cover assembly.

2. Place the top cover assembly top down on a cloth or other soft surface.

3. Disconnect the paper sense cable connector from the recorder adapter board.

4. Turn over the top cover and remove the screw holding the paper sensor to the RFI bracket


Replacing the Paper Sensor Assembly

The procedure for replacing the paper sensor is a reversal of the removal procedure. Ensure that the paper sensor cable is properly connected to the recorder adapter board.


► Company Profile

We cater to all biomedical facility needs - service, repair, rental, and purchase and sale of equipment. Safe, reliable and compliant medical device services are our primary goals. We pride ourselves on the individualized attention we provide each client, catering to all aspects of their biomedical facility. Our team of experienced biomedical technicians are trained, certified and equipped with the latest technology and tools to repair and maintain a wide variety of patient care and diagnostic equipment.

Our technicians ensure that all equipment leaving our warehouse is meticulously tested and certified for safety and compliance. We understand the importance of device safety and performance in today's healthcare environment. Providing the most professional customer service and adhering to the highest quality standards is our mission.


We seek for providing perfect service to the satisfaction of customers. Thank you for your time!

If you are interested in PHiliphh FM20 Fetal Monitor Paper Sensor Board Assembly M1353 66522 M2703-60003, welcome to place your orders. Early order early shipment!

China PHiliph FM20 Fetal Monitor Paper Sensor Board Assembly M1353 66522 M2703-60003 supplier

PHiliph FM20 Fetal Monitor Paper Sensor Board Assembly M1353 66522 M2703-60003

Inquiry Cart 0