Add to Cart
The product is JC 23V-100W halogen lamp Maquet surgical shadowless lamp ECL-0001 bulb
Shadowless lamps use multiple light sources to illuminate from
different angles, combined with high reflectivity materials and
adjustable brightness design, to eliminate shadows in the surgical
area, provide uniform and clear lighting, and have heat dissipation
and sterile design to ensure surgical accuracy and safety.
Shadowless lamps are widely used in the medical field, including
surgery, dentistry, otolaryngology, emergency, cosmetic surgery,
laboratory research, and veterinary medicine, providing clear,
shadowless high brightness lighting for people in various fields,
supporting precise and safe medical operations.
a. Check the power fuse for damage/looseness caused by poor
contact, and everything is normal after investigation.
b. The power module at the base of the power module detection
equipment was checked for any damage to electrical components, and
no abnormalities were found after inspection.
c. The controller module was inspected and disassembled to check
the circuit module at the control operation board for
looseness/damage. After inspection, no abnormalities were found.
d. Firstly, for the inspection of the lamp head module, remove the
lamp head and check if there is any burnt out phenomenon in the
lamp beads. After inspection, the lamp beads are found to be
normal.
Secondly, whether there are any abnormalities in the circuit
modules and circuits; After careful inspection, it was found that
there were burn marks on the main light socket.
If the problem cannot be solved as mentioned above, it needs to be
dealt with without replacement parts. After the main lamp holder is
rust removed (without replacement parts), if the fault still
occurs, the lamp holder and related circuits should be replaced.
Firstly, use rust remover to remove rust and check if the fault
still occurs during later use. If the fault does not occur, there
is no need to replace it.
Type: | JC 23V-100W halogen lamp Maquet surgical shadowless lamp ECL-0001 bulb |
Warranty period: | 90 Days |
Shipping Date: | 3-5 Working Days |
MODLE: | 23V-100W halogen lamp ECL-0001 bulb |
Storage temperature: | -20-70°c |
MOQ: | 1 PCS |
COLOR: | Metal and glass |
Delivery Date: | 3 - 5 Working Days |
Package | Carton |
The company has been professionally operating various brands of original medical monitoring instruments since 2015. The company has many professional senior engineers and has sufficient ability to undertake various medical equipment maintenance work. Covering motherboards, sensors, probes, modules, connectors, shells, etc. for various medical monitoring equipment.
We default to express delivery by DHL,Fedex, We will contact you as soon as possible when sending express delivery. If you have any questions after receiving the express delivery, please feel free to contact us at any time.