Add to Cart
Thank you for browsing the TEPHN NT-1000P 12V30W suitable for
Neusoft fully automatic biochemical analyzer light source bulb. The
light source uses a halogen tungsten lamp as the light source, with
up to 16 wavelengths and an effective usage time of not less than
2000 hours. It also uses air cooling and heat dissipation to extend
its service life. When the effective time is exceeded or the gain
value of the 340nm wavelength is 4.0, it is necessary to replace
the light source lamp in a timely manner to ensure the accuracy and
stability of the test results. The advantages of this light source
are: it can generate strong continuous spectra in some ultraviolet
regions and the entire visible light range, with low noise and
small drift;
Disadvantages: Short lifespan and time limit.
When the usage time of the light source lamp approaches or reaches
its service life, the emitted light intensity will experience
energy attenuation, which will affect the detection results.
The light source lamp, as the beam provider in the biochemical
analyzer, is crucial in the entire reaction system. Most
biochemical analyzers use halogen lamps as the light source, with a
working wavelength of 340-804nm. For most enzyme rate based
biochemical detection projects, the quality control value changes
in a regular manner with the performance changes of the light
source lamp. After abnormal curves or results appear, confirmation
is needed. At the same time, we need to regularly maintain and
calibrate the light source of the biochemical analyzer.
It can be divided into several steps:
1. Maintenance - Routine maintenance - Automatic calibration of
photometer
2. Cleaning the color comparison cup: It is necessary to perform a
daily cleaning and maintenance of the color comparison cup after
turning it on - regular maintenance - cleaning the color comparison
cup
3. Cup blank check: Maintenance - Routine maintenance - If the 1 #
colorimetric cup displays red or yellow, the photometer needs to be
automatically calibrated again.
If in contact with sampling mechanisms, reagent dispensing
mechanisms, stirring mechanisms, etc., it may cause injury or
infection.
When conducting maintenance checks, please be sure to wear
protective tools and follow the order in the user manual. When
samples, reagents, etc. adhere to the body, please immediately
rinse and disinfect with water, and accept a doctor's diagnosis
Type: | TEPHN NT-1000P 12V30W suitable for Neusoft fully automatic biochemical analyzer light source bulb |
Warranty period: | 90 Days |
Shipping Date: | 3-5 Working Days |
MODLE: | TEPHN NT-1000P 12V30W |
Storage temperature: | -20-70°c |
MOQ: | 1 PCS |
COLOR: | Metal and glass |
Delivery Date: | 3 - 5 Working Days |
Package | Carton |
The company has been professionally operating various brands of original medical monitoring instruments since 2015. The company has many professional senior engineers and has sufficient ability to undertake various medical equipment maintenance work. Covering motherboards, sensors, probes, modules, connectors, shells, etc. for various medical monitoring equipment.
We default to express delivery by DHL,Fedex, We will contact you as soon as possible when sending express delivery. If you have any questions after receiving the express delivery, please feel free to contact us at any time.