3 In 1 AC DC Digital Volt Ohm Meter HFE NVC Capacitor Meter Tester
AC/DC Ammeter Voltmeter Ohm Voltage Meter HFE NVC Capacitance Tester
Non - contact voltage measurement(NCV)
The function range switch is set to NCV range position. At this
time, the instrument displays "NCV" and "EF" symbol, and the top of
the instrument is close to the object to be tested such as mains,
power switch, socket and so on. When the detected voltage is
greater than 110V (AC RMS), the instrument NCV indicator will be
lit and the buzzer will sound an alarm.
Technical indicators
Ambient temperature: 23 ± 5 ℃ Relative humidity: <75%
3.2.1 True rms zero input
3.2.1.1 The measurement of sine wave signal, the use of true RMS
measurement method than the traditional average response method
error is small.
3.2.1.2 True RMS The instrument can accurately measure the sine
wave signal, but if it is on the AC function, the instrument may
display 1 to 50 if the measured signal is not input (shorted at the
AC voltage range and the input terminal) Word between a reading.
These deviations are normal. Within the specified measuring range,
they do not affect the accuracy of the multimeter to measure
alternating current.
3.2.1.3 True RMS requires that the input signal reach a certain
level for measurement. Therefore, the AC voltage and current range
is specified between 2% and 100% of full scale.
3.2.2 DC voltage
Range | Resolution | Accuracy |
6V | 0.001V | ±(0.5% of reading + 5 words) |
60V | 0.01V | |
600V | 0.1V | ±(0.8% of reading + 5 words) |
- Input impedance: 10MΩ
- Maximum input voltage: 600V DC or 600V AC (rms)
Note:
In the small voltage range, the pen is not connected to the circuit
under test, the instrument may have beating the reading,
This is normal, this is because the instrument caused by high
sensitivity, does not affect the actual measurement results.
3.2.3 AC voltage
Range | Resolution | Accuracy |
6V | 0.001V | ±(0.8% of reading + 3words) |
60V | 0.01V | |
600V | 0.1V | ±(1.0% of reading + 10words) |
- Input impedance: 10MΩ
- Maximum input voltage: 600V AC (RMS) or 600V DC
- Frequency range: 40 ~ 1000Hz True RMS
Note:
In the small voltage range, the pen is not connected to the circuit
under test, the instrument may have beating the reading,
This is normal, this is because the instrument caused by high
sensitivity, does not affect the actual measurement results.
3.2.4 Alternating current
Range | Resolution | Accuracy |
600.0uA | 0.1uA | ±(1.5% of reading +10words) |
6000uA | 1uA | |
60.00mA | 0.01mA | |
600.0mA | 0.1mA | |
6A | 0.001A | ±(2.5% of reading +10words) |
10A | 0.01A |
- Overload protection: mA range fuse (self-recovery fuse 200mA /
250V)
10A range without insurance
- Maximum input current: mA: 600mA DC or AC rms;
10A: 10A DC or AC rms;
When the measured current is greater than 5A, the continuous
measurement time is not longer than 15 seconds, after the
measurement to stop the current measurement for more than 1 minute.
- Frequency response: 40Hz ~ 1000Hz True RMS.
3.2.5 DC
Range | Resolution | Accuracy |
600.0uA | 0.1uA | ±(1.0% of reading +5words) |
6000uA | 1uA | |
60.00mA | 0.01mA | |
600.0mA | 0.1mA | |
6A | 0.001A | ±(2.0% of reading +10words) |
10A | 0.01A |
- Overload protection: mA range fuse (self-recovery fuse 200mA /
250V)
10A range without insurance
- Maximum input current: mA: 600mA DC or AC rms;
10A: 10A DC or AC rms;
When the measured current is greater than 5A, the continuous
measurement time is not longer than 15 seconds, after the
measurement to stop the current measurement for more than 1 minute.
3.2.6 Frequency
Range | Resolution | Accuracy |
9.999Hz | 0.001Hz | ±(1.0% of reading +3words) |
99.99Hz | 0.01Hz | |
999.9Hz | 0.1Hz | |
9.999kHz | 0.001kHz | |
99.99KHZ | 0.01kHZ | |
999.9KHZ | 0.1KHZ | |
9.999MHZ | 0.001MHZ |
- Overload protection: 250V DC or AC (rms)
- Input voltage range: 600mV ~ 10V AC (with the measured frequency
increases, the input voltage should also increase)
3.2.7 Duty cycle
Range | Resolution | Accuracy |
0.1 – 99.9% | 0.1% | ± 3.0% |
- Frequency response: 10 ~ 10MHz
- Input voltage range: 600mV ~ 10VAC (with the measured frequency increases, the input voltage should also increase)
- Maximum input voltage: 250V AC (rms)
Resistance
Range | Resolution | Accuracy |
600W | 0.1W | ±(0.8% of reading +3words) |
6kW | 0.001kW | |
60kW | 0.01kW | |
600kW | 0.1kW | |
6MW | 0.001MW | ±(1.2% of reading +3words) |
60MW | 0.01MW |
- Open circuit voltage: approx. 1.0V
- Overload protection: 250V DC or AC (rms)
Transistor test
Range | Description | Test Conditions |
hFE | The display reads the approximate value of hFE(0~1000) | Base current of about 10uA; Vce about 2.8V |
Capacitance
Range | Resolution | Accuracy |
9.999nF | 0.001nF | ±(4.0% of reading +3words) |
99.99nF | 0.01nF | |
999.9nF | 0.1nF | |
9.999mF | 0.001mF | |
99.99mF | 0.01mF | |
999.9mF | 0.1mF | |
9.999mF | 0.001mF | ±(5.0% of reading +3words) |
99.99mF | 0.01mF | Not calibrated |
- Overload protection: 250V DC or AC (rms)