Premium ULPA Filter HD LED Display BSC Biological Safety Cabinet
BSC Biosafety Cabinet Features:
1. The biological safety cabinet has four negative pressures,
namely: the left and right sides of the entire device, the rear,
and the bottom wall are all negative pressure air ducts, so that
the working area and the external environment form a double working
air curtain and the cabinet to maintain the isolation. The area is
surrounded by negative pressure to ensure that the sample does not
leak.
2. The cabinet body part adopts 1.5mm thick cold-rolled steel plate
to enhance the structural strength. The surface of the cabinet body
is electrostatically sprayed to make the whole device more stable.
3. The control panel adopts light touch switch, humanized design,
beautiful appearance of the machine and convenient operation.
4. The working area inside the workbench and the equipment is
high-quality 304# stainless steel, which is corrosion-resistant.
5. Ultra high efficiency filter for particles, smoke and
microorganisms in the air filter. Various types of biological
safety cabinets use imported ultra-high efficiency filters (ULPA)
with a filtration efficiency of over 99.995% (0.1-0.2μmDOP), and
the cleanliness of the working space can reach 100. Equipped with a
dual-channel digital pressure monitoring and electronic alarm
system, real-time dynamic monitoring of the pressure loss of
exhaust and air filters, downward flow and inflow air flow in the
working area, two-way detection,
6. The opening height of the safety cabinet window exceeds the
height specified by the front window operation port, sound and
light alarms, and the interlock system is activated. When the
opening height returns to the specified opening height, the alarm
interlock system will automatically release.
Model | BSC-1300B2 |
Clean Grade | HEPA:ISO 5 calss ( Class100 ) |
ULPA:ISO 4 Class ( Class10 ) | |
HEPA Filter | HEPA:≥99.995% , @0.3μm |
ULPA:≥99.999% , @0.12μm | |
Air tightness | 99.99≤E≤99.999 |
Overall Dimensions | 1500×800×2120mm |
Working Space | 1300×700×600 mm |
Downflow Velocity | 0.35m/s |
Forward intake wind speed | 0.25m/s±0.025m/s |
Installation of BSC Biosafety Cabinet:
1. The biological safety cabinet shall not be placed on its side, impacted, or collided during transportation, and shall not be directly attacked by rain, snow, or sun exposure.
2. The working environment of the biological safety cabinet is 10~30℃ and the relative humidity is <75%.
3. The equipment should be installed on a horizontal surface that cannot be moved.
4. The equipment must be installed close to a fixed power socket.
5. In order to prevent airflow interference, it is required that the equipment should not be installed in the passages of people, and the operating port of the sliding front window of the biological safety cabinet should not be directly facing the doors and windows of the laboratory or too close to the doors and windows. Places that may disturb the airflow.