Insulation Resistance Meter ZC-11d Safety Testing Meter
ZC11D
Use Instruction of ZC11D Type Insulation Resistance Meter
I. Purpose
,,,,.
The instrument is used to measure the insulation resistances in
various motors, cables, transformers, telecom elements, household
electrical appliances and other electrical devices.
ZC11D
ZC11D type meter is mainly applicable to measurement of insulation
of resistance wire withstanding voltage.
II. Specification and performance
1. Using condition: temperature (-25ºC-+40)ºC, relative humidity
not more than 80%
2. Rated rotation rate of rocking handle: 150r/min
:
3. Length of graduated arc: about 80mm
4. Outline size: axbxh (215x135x130)mm
5. Weight: about 2.5KG
6. Influence from incline: when the meter inclines from horizontal
position to any direction by 5 º, the allowable change volume of
the meter is 50% of the grade index.
7. Influence from external magnetic field: when strength of
external magnetic field is 5 º, the allowable change volume of the
meter is 100% of the grade index.
8. Specification: measurement range, rated output voltage,
accuracy, insulation resistance withstanding voltage, etc. are
shown in the table.
Model | Rated voltage | (MΩ) Measurement range (MΩ) | (%) Grade accuracy (%) | (kv) Withstanding voltage (kv) | |
(V) Voltage (V) | (%) Accuracy (%) | ||||
SBE11D-1 | 100 | 10 | 0-500 | 10 | 1 |
SBE11D-2 | 250 | 0-1000 | 10 | 1 | |
SBE11D-3 | 500 | 0-2000 | 10 | 1 | |
SBE11D-4 | 1000 | 0-5000 | 10 | ||
SBE11D-5 | 2500 | 0-10000 | 20 | 3 | |
SBE11D-6 | 100 | 0-20 | 10 | 1 | |
SBE11D-7 | 250 | 0-50 | 10 | 1 | |
SBE11D-8 | 500 | 0-100 | 10 | 1 | |
SBE11D-9 | 50 | 0-200 | 10 | 1 | |
SBE11D-10 | 2500 | 0-2500 | 20 | 3 |
III. Operation and instruction
,.
1. The meter shall be kept away from magnetic field during
operation and placed in horizontal position.
"
2. During insulation measurement, both ends of the measured
subjects can be respectively connected to the binding posts of
"grounding" and "line" and the resistance value can be measured by
rotating the handle. See figure I.
Figure I
.
3. For grounding measurement, the measured end is connected to
"line" and the grounding line with good condition is connected to
"grounding" post. See figure II.
Figure II
4. For measurement of insulation of cable core to cable shield,
besides connecting the two measured ends to the wiring posts of
"grounding" and "line", the liner insulation material between cable
core and shield should be connected to "shield" to eliminate the
measurement error caused by surface electric leakage. See figure
III.
Figure III
5. Stranded wire can't be used for connection between the wiring
post and the measured subject, which should be connected separately
so that reading will not be influenced by poor insulation of the
stranded wire.
6. The meter must not be used to measure the devices close to
lightning or equipped with high voltage conductors. The meter can
only by operated when the device is electrically neutral and will
not be charged by being induced by other power sources.
7. After measurement, the measured subject must be discharged
sufficiently to assure safety of persons and devices.
8. Rotation rate of the handle should grow faster gradually from
slow rate; fierce rocking should be forbidden when the pointer is
at zero to prevent coils from damaging.
9. The fierce and long-term vibration should be avoided so that
accuracy will not be influenced by the damaged axis end of the
meter head and the gems.
10. The meter should be kept in a fixed cabinet during idle period
and the ambient temperature should not be too low or two high; the
meter must not be placed on the dirty and humid ground and or near
the corrosive air (e.g. steam of acid, alkali, etc.)
IV. Electrical schematic diagram
G-generator
E- earthing
L - line
G - shielding
D -silicon diode
C - capacitor
Ri -current-limiting resistance of current coil
Rv - current-limiting resistance of voltage coil
Li - current coil
Lv - Balance coil
The meter applies to international standard of GB/T7676-1998
V. Package set
1. SBE11D 1
1.SBE11D type insulation resistance meter 1
2. Check bar 1
3. Use instruction 1
4. Quality certificate 1