Add to Cart
YD5010 Plus Grating Spectrophotometer
New Model YD5010 Plus Grating Spectrophotometer Densitometer with
45/0(45 ring-shaped illumination, 0 degree viewing angle), comply
to standard ISO 5-4,CIE No.15. It is suitable for using in ink
printing, in Film Processing, Textile Printing and Dyeing, Plastic
Electronics and Other Industries for Color Measurement and Quality
Control;especially suitable for the measurement and quality control
of optical density and dot enlargement in ink printing; and it is
able to customize single aperture Φ2mm,Φ4mm,Φ8mm.
Under the 45/0 geometric optical illumination and the testing
conditions of M O, M 1,M 2 , M 3 stipulated by ISO 13655 standard,
the instrument can accurately measure the reflectance data of
samples. Under multiple color spaces, it can accurately measure
various printing density indexes, color difference formulas and
color indexes, and can meet the user's regular testing of various
parameters.
1.Perfect combination of the beautiful appearance and the ergonomic structure design;
2.Combined LED light sources with long life and low power consumption, including UV light;
3.Optional apertures: Φ2/4/8mm, adapt to more samples;
4. Accurately measure reflectance spectrum, CMYK density and Lab value of the sample;
5.High-configuration electronic hardware: 3.5-inch TFT true-color screen, capacitive touch screen, concave grating, 256-pixel dual-array CMOS image sensor, etc.;
6.Two standard observer angles: 2/10, multiple light source modes and color systems;
7.USB mode is widely useful;
8.Large-capacity storage space, over 10,000 test data;
9.Especially suitable for process control and quality control of printing plants;
10. PC software has powerful function expansion.
Model | YD5010 Plus |
Measurement Geometry | 45/0(45 ring-shaped illumination, 0 degree viewing angle); ISO 5-4,CIE No.15 |
Light Source | Combined LED Light, UV Light |
Spectral separation device | Concave Grating |
Detector | 256 Image Element Double Array CMOS Image Sensor |
Wavelength Range | 400~700nm |
Wavelength Pitch | 10nm |
Half Bandwidth | 10nm |
Measurement Conditions | Compliance with ISO 13655 measurement conditions; M0 (CIE Light Soure A) M1 (CIE Light Soure D50) M2 (Excluding UV light source) M3 (M2+Polarized light filter) |
Measurement aperture | Customized one aperture:Φ2mm,Φ4mm,Φ8mm optional |
Color Spaces | CIE LAB,XYZ,Yxy,LCh |
Color Difference Formula | ΔE*ab,ΔE*94,ΔE*00 |
Observer Angle | 2°/10° |
Illuminants | A,C,D50,D55,D65,D75,F2,F7,F11,F12 |
Measurement Time | About 1.5s |
Repeatability | Density: Within 0.01 D |
Inter-instrument Error | Within ΔE*ab 0.2 (Average for 12 BCRA Series II color tiles) Except M3 |
Measurement mode | Single Measurement, Average Measurement(2-99) |
Size(L*W*H) | 184*77*105mm |
Weight | About 600g |