Add to Cart
SH105B automatic closed cup flash point tester is designed and manufactured as per standards GB/T261Determination of Flash Point - Pesky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products in the temperature range from 40℃ to 360℃.
Technical features
1. It adopts LCD screen to display. It can preset the parameters of expected flash point,sample mark number, atmospheric pressure, test date,etc.It has menu to prompt and input function of guide type.
2. It adopts simulation tracking display the function curve of temperature rising and test time.It has prompt functions mis-operation,test date and test time.
3. It is equipped with standard RS-323,485 computer port.The lower computer can save 100 groups of historical data. The test data can be saved for a long time, transmitted and modified if the instrument being connected with computer. Pensky-Martens Closed Cup Tester
4. It can correct the influence of atmospheric pressure and calculate the correction automatically.
5. Differential detection.Automatically correct the systematic deviation.
6. Automatically open the lid,ignite,detect and print test data.The test arm lifting up and down automatically.
7. It adopts electric ignition to fire light the gas flame.It will light only need to press the button.
8. Humanized design,beautiful and safe. Easy to operate..
Technical specifications and parameters
1. Power :AC(220±10%)V, 50Hz
2. Temperature measurement:Range: room temperature~300℃
Repeatability: ≤3℃
Reproducibility:≤6℃ Pensky-Martens Closed Cup Tester
Accuracy:0.1℃
3. Heating rate:Procedure A:(5~6)℃/min
Procedure B:(1~1.6)℃/min
Automatic control and manually adjustable.
4. Stirring rate:(1) Procedure A: (90~120)RPM
(2) Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5. Igniting mode:Electric ignition. Gas flame diameter:3.2mm~4.8mm
6. Working condition:ambient temperature:(10~40)℃
relative humidity:≤80%
7. Total power :≤500W
8. Overall dimension: 520mm×360mm×310mm
Temperature measurement Range | room temperature~300℃ |
Power | AC(220±10%)V, 50Hz |
Heating rate:Procedure A | (5~6)℃/min |
Heating Procedure B | (1~1.6)℃/min |
Igniting mode | Electric ignition |
Total power | ≤500W |
Overall dimension | 520mm×360mm×310mm |
OUR service
After-sales service commitment, our company to sell the product after-sales service commitment: warranty period of one year, long-term technical support outside the warranty period. During the warranty period, the seller shall bear the cost of the replacement of the equipment, the freight and maintenance of the instrument; outside the warranty period, the seller shall waive the maintenance fee and charge only the cost of the basic materials and freight charges when the instrument is malfunctioning and the seller shall not charge the maintenance fee during the warranty period.
Contact us
Shandong Shengtai Instrument Co.,ltd
ADD:6-8 time base NO.15 lanxiang road tianqiao Area jinan city
TEL:+86 0531-85919987 85929987
FAX:+86 0531-85957987
Wechat:+86 13370527859
Whatsapp/facebook:+86 13370527859
QQ:3206156217
Email:3206156217@qq.com
Web:www.labtool.net/en