GC Introduction
The intelligent Gas Chromatograph GC-5400 incorporates advanced modern manufacturing technologies with years of experience in research and development. It is manufactured under a quality system conforming to international standards. The multi functional, easy-to-use and remotely controllable gas chromatograph provides superior performance for all applications.
Application fields
GC-5400 gas chromatograph plays a critical role in quality control of chemistry, medicine and food industries, and also applicable to quarantine, environment protection, quality supervision, oil, petrochemical industry, forensic analysis, agriculture and commodity inspection.
Performance characteristics
With adoption of advanced control technology, the microcomputer
temperature control system is highly precise (less than ±0.05℃),
reliable and anti-interfering. The system is divided into 5
independent heated zones with maximum temperature up to 400℃.
Settable ultimate temperature and overheat protection mechanism
ensure a safe operation of the instrument.
Control and operating parameters can be set on the logical and
easy-to-use keypad interface (including detector operating
parameters). It has functions of self diagnosis, power-off
protection, 10 method files saving and invoking, 2 controls of
external events, timing start/stop of temperature control, time
programming, detector selection, setting and display of range,
polarity and current. The set value, actual value, retention and
analysis time of each heated zone can be fully displayed.
The large column compartment allows for simultaneous accommodation of capillary column and double packed columns. It is equipped with automated back door mechanism, a rapid heating/cooling device (lowering temperature from 350℃ to 50℃ in 7 minutes), to realize quasi room temperature control. It supports 5 compartment ramps with 6 plateaus.
The carrier gas pneumatics is designed with constant pressure valve plus constant flow valve for dual stability. Constant pressure valve plus needle valve are used in the air and Hydrogen pneumatics for stable gas regulation. Both valves are in form of digital knobs, which delivers high precision, outstanding repeatability, good reliability and intuitive flow rate adjustment.
Up to three kinds of detectors, Flame Ionization Detector (FID),
Thermo Conductivity Detector (TCD) and Electron Capture Detector
(ECD), are provided for GC-5400. Different combinations of
detectors, in series or parallel operation, can meet different
application requirements.
The novel FID has lower limit of detection, with ejector nozzle and
ions collector easily disassembled or cleaned. The fixed and
accurate installation structure of the ejector ensures every
instrument has consistent performance.
The high sensitive TCD adopts the high resistant rhenium-tungsten
wire, with twofold electronic amplification circuit optional,
resulting in maximum sensitivity up to 5000mV·mL/mg. Its unique air
insulation structure brings TCD higher stability of performance.
Sensitivity and current selections of pulse modulation ECD constant
current source are computerized.
Technical specifications
Temperature Control | |
Temperature range | 7°C ~ 400°C above room temperature, increment 1°C |
Temperature control objects | Column compartment, detector, thermal conductivity cell, injector, accessories |
Temperature ramps | 5-stage |
Temperature ramp rate | 0. 1°C ~ 40°C/min, increment 0. 1°C |
Constant Temperature Retention time | 0 ~ 655 min, increment 1 min |
Hydrogen Flame Ionization Detector (FID) | |
Sensitivity | M ≤ 1 ×10-11 g/s(C16) |
Best testing results | M ≤ 8 ×10-12 g/s(C16) |
Noise | ≤5×10-14 A |
Drift | ≤6×10-13 A/h |
Linear dynamic range | ≥ 106 |