Air Measurement Vortex Gas Flow Meter With RS485 4 - 20 mA Pulse Output
1. Introduction
LUGBML is based on Von Karman’s Vortex Principle, which adopts piezoelectric crystal as testing components. It is a new kind of stress detecting vortex flow meter. It has the advantage of wide range ratio, high accuracy, low pressure loss, good medium generality, pulse output proportional to flow, convenient to connect with computer and etc. As the detector probe of sensor and the vortex generating body are installed separately, and the high temperature resistant piezoelectric crystal doesn’t contact with medium, so the instrument is of simple structure, good generality and high stability.
LUGBML series Vortex Flow Meter sensor can be used to detect and measure the flow of all kinds of gas, liquid and steam.
LUGBML series Vortex Flow Meter can be used to match with our company’s ML6000 series Velocity Flowmeter. It can also match with a computer and sensors of temperature, pressure, or density to compose high accuracy mass flow or heat flow detecting and measuring system.
2. Features
3. Specifications
1) Measuring medium: Steam,gas, liquid(Avoid of multiple- direction fluid)
2) Accuracy:Liquid:1.0% Gas:1.5%
3) Repeatability:Liquid 0.33% Gas:0.5%
4) Range:Liquid:0.7m/s~7m/s Gas:7m/s~40m/s
5) Working pressure:Standard 1.6MPa,2.5~4.0 MPa (on request)
6) Fluid temperature:-40℃~250℃(Common type);100℃~350℃(High-temperature type)
7) Body material:304 stainless steel(SS316 is optional)
8) Output signal:Pulse(three wires,low level≤1V;High level≧4V);4~20mA;RS485
9) Power supply:24VDC(12VDC),Lithium battery
10) Ambient temperature : -35 ℃ ~ 60 ℃ ( without LCD ); -5℃~60℃(with LCD)
11) Relative humidity:5~95%
12) Protection level:IP65
13) Electrical connection:M20*1.5
4. Dimensional Sketche
DN | A(mm) | B(mm) | C(mm) |
25 | 65 | 65 | 368 |
32 | 66 | 65 | 374 |
40 | 80 | 76 | 382 |
50 | 80 | 89 | 388 |
65 | 93 | 102 | 402 |
80 | 100 | 114 | 408 |
100 | 126 | 136 | 435 |
125 | 146 | 160 | 463 |
150 | 166 | 182 | 489 |
200 | 196 | 248 | 543 |
250 | 210 | 282 | 596 |
300 | 240 | 334 | 648 |
5. Design & Installation
It is important to install meter,if not installed well, then would
affect precision, use-life and damage.
1. To avoid strong current, high frequency and powerful switch set,
power supply of meter shall be avoided to near by these equipments.
2. To avoid high-Temp & radiation source. if have to install
it, need heat insulation & ventilated measure.
3. To avoid high-Temp & enchant gas, if have to install it,
need ventilated measure.
4. Vortex flow meter shall be avoided to install on shaking part of
pipeline. If have to install on it,shall add clamp device and
vibration pad which located on 2D to enhance shake proof . meter
has better to installed indoors, pay attention to waterproof when
installing meter outdoors,special notice the joint, make cable
conductor to U shape to avoid water get into the amplifier body
Around installing place shall save enough space,so that install
connection line and maintenance routine.
6. Details