Add to Cart
IEC 60601-2-52 Figure CC.1 Wedge Tool For Test Medical Bed In Non-Moving Parts
Standard & Clauses:
IEC60601-2-52:2015figure CC.1
Specimen Range, Application, Principle, Structure and Use Environment:
Specimen and Limitations: Medical Bed
Test purpose: used to evaluate the risk of V-shaped openings
getting stuck in areas B and C between the edge of the medical bed,
the headboard/footboard and the mattress support platform, and to
verify whether the gaps meet the standard requirements.
Test principle: Use a wedge-shaped tool to measure at various
positions required by the standard, and judge according to the
standard.
Equipment structure and features: Made of aluminum
Weight: 3.34kg
Features:
To reduce the risk of neck entrapment, openings between elements in
the MEDICAL BED systemabove the MATTRESS SUPPORT PLATFORM should be
small enough to prevent passage of asmall neck. Additionally to
reduce neck entrapment with regards to wedging of the neck,angles
for V-shaped entry into the openings should be greater than 75°.
The 75° anglereflects an understanding that at some point, a narrow
angle creates a V-shaped space orgeometry that could be conducive
to entrapment, The 75° angle was chosen by reviewing thetools to
assess entrapment risk for swimming pool equipment (see EN 13451-1)
and makingchanges suitable to application on MEDIcAL BEDs by
increasing the wedging minimum angle o60° to 75°. The 75° angle
reduces the RiSK of PATlENT entrapment through hooking of the
neckand head in the V-shaped opening (See Figure AA.13 and Figure
AA.15). The wider angleallows the PATiENT to pull his head free
from a vertical opening with less effort. This isessential when
considering a weak or confused PATIENT.
To measure V-shaped openings in relation to areas B and C a test
method and wedge tool(see Figure CC,1) has been developed, The tool
is made from aluminium and weighs 3,34 kgrepresenting the mass of a
small aduits head and neck (the 3,34 kg was taken from Air
ForceResearch laboratory Design and Development of
Anthropometrically Correct Head Forms forJoint Strike Fiahter
Eiection Seat Testing-February 2005), the wedge angle is 75° and
has a60 mm diameter cylindrical end representing the small adult
neck.