Fluorescent qPCR Cat Stomatitis Multiple Pathogens ( FHV / FCV / FeLV / FIV ) DNA detection kit
Intend for Use:
This kit is suitable for the specific detection of feline herpes
virus (FHV), feline calicivirus (FCV), feline leukaemia virus
(FeLV) and feline AIDS virus (FIV) in cat eye, nose and mouth
secretions and EDTA whole blood samples.
Principle:
Under the action of high-efficiency reverse transcriptase, cDNA
strands complementary to RNA template were synthesized with PRIMER
as starting point and RNA template as template. Choose cat herpes
virus, cat goblet virus, feline leukemia virus, HIV gene
conservative design specific probe, the probe can with primer
amplification area in the middle of a specific binding DNA
template, in the process of PCR extension, the circumscribed Taq
polymerase enzymes from the 5 'end of fluorescent groups on the
probe will be cut down, make it free in the reaction system, In
this way, the 3 '-end quenched group is separated, and the
fluorescence is emitted, and the nucleic acid detection of feline
herpes virus, feline calicivirus, feline leukaemia virus and feline
AIDS virus is realized in the completely closed reaction system by
fluorescence PCR instrument.
Component:
Composition | Quantity |
FHV/FCV PCR liquid | 8 tube |
FeLV/FIV PCR liquid | 8 tube |
Positive control | 1 tube |
Negative control | 1tube |
FHV/FCV PCR Reaction reagent:
FAM channel (FPV) | VIC/HEX channel(FCoV) | Result interpretation | |
1 | - | - | Results were reported negative |
2 | + | - | Results were reported FHV positive |
3 | - | + | Results were reported FCV positive |
4 | + | + | Results were reported FHV,FCV positive |
Note: "+" means positive, "-" means negative
FeLV/FIV PCR PCR Reaction reagent:
FAM channel (TF) | VIC/HEX channel(GD) | Result interpretation | |
1 | - | - | Results were reported negative |
2 | + | - | Results were reported FeLV positive |
3 | - | + | Results were reported FIV positive |
4 | + | + | Results were reported |
Product performance index:
1. Accuracy
The positive reference material of testing enterprises were all positive.
2. Minimum detection limit: The detection limit of this kit is 10 copies/μL.
3. Precision: Coefficient of variation (CV,%) of intra-batch
precision Ct value is ≤5%.